Multi‐site feasibility and reproducibility study on UTE 3D phosphorous MRSI using novel rosette trajectory ( PETALUTE )
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30640 ↗At a glance
Multizentrische Studie in Magnetic Resonance in Medicine, 2025. Written by Seyma Alcicek and 7 co-authors. Institution: Goethe University Frankfurt, University Hospital, Institute of Neuroradiology, Frankfurt am Main, Germany. Major MEDLINE topics: Imaging, Three-Dimensional, Brain, Magnetic Resonance Imaging. Cited 1 times so far (Crossref); 42 references. The full text is freely available (open access, CC BY 4.0). Funded by: Indiana Clinical and Translational Sciences Institute, National Institutes of Health (USA), Else Kröner-Fresenius-Stiftung (EKFS).
Abstract (original language)
Purpose
This study aims (1) to implement a robust acquisition, fully automated reconstruction, and processing pipeline using a novel rosette k‐space pattern for UTE 31 P 3D MRSI and (2) to evaluate the clinical applicability and reproducibility at different experimental setups.
Methods
A multi‐center feasibility/reproducibility study was conducted for the novel UTE 31 P 3D MRSI sequence with rosette petal trajectory (PETALUTE) at three institutions with different experimental setups (Siemens Prisma with volume head coil or surface coil, Siemens Biograph mMR with volume head coil). Five healthy subjects at each site were measured with an acquisition delay of 65 μs and a final resolution of 10 × 10 × 10 mm 3 in 9 min. The measurement was repeated three times and averaged for the spectral analysis using the LCModel package. The potential for acceleration was assessed using compressed sensing on retrospectively undersampled data. Reproducibility at each site was evaluated using the inter‐subject coefficient of variance.
Results
This novel acquisition and advanced processing techniques yielded high‐quality spectra and enabled the detection of the critical brain metabolites at three different sites with different hardware specifications. In vivo, feasibility with an acceleration factor of 4 in 6.75 min resulted in a mean Cramér‐Rao lower bounds below 20% for phosphocreatine (PCr), adenosine triphosphate (ATP), phosphomonoesters (PME), and a mean coefficient of variation for ATP/PCr below 20%.
Conclusion
We demonstrated that UTE 31 P 3D rosette MRSI acquisition, combined with compressed sensing and LCModel analysis, allows clinically feasible, robust, high‐resolution 31 P MRSI to be acquired at clinical setups.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Goethe University Frankfurt, University Hospital, Institute of Neuroradiology, Frankfurt am Main, Germany.
- University Cancer Center Frankfurt (UCT), Frankfurt am Main, Germany.
- Frankfurt Cancer Institute (FCI), Frankfurt am Main, Germany.
- Department of Clinical Engineering, Haukeland University Hospital, Bergen, Norway.
- Department of Biological and Medical Psychology, University of Bergen, Bergen, Norway.
- Radiology and biomedical imaging, University of California, San Francisco, California, USA.
ORCID profiles: Seyma Alcicek, Xin Shen, Mark Chiew, Uzay E Emir
Funding
- Indiana Clinical and Translational Sciences Institute
- National Institutes of Health (USA)
- Else Kröner-Fresenius-Stiftung (EKFS)
- Wellcome Trust
- NCATS NIH HHS
- Mildred Scheel Career Center (Deutsche Krebshilfe)
- Deutsche Krebshilfe
Cite
APA
Alcicek, S., Craven, A. R., Shen, X., Chiew, M., Ozen, A., Sawiak, S., Pilatus, U., & Emir, U. E. (2025). Multi‐site feasibility and reproducibility study on UTE 3D phosphorous MRSI using novel rosette trajectory ( PETALUTE ). Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30640
Vancouver
Alcicek S, Craven AR, Shen X, Chiew M, Ozen A, Sawiak S, et al. Multi‐site feasibility and reproducibility study on UTE 3D phosphorous MRSI using novel rosette trajectory ( PETALUTE ). Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30640
BibTeX
@article{Alcicek2025,
author = {Alcicek, Seyma and Craven, Alexander R and Shen, Xin and Chiew, Mark and Ozen, Ali and Sawiak, Stephen and Pilatus, Ulrich and Emir, Uzay E},
title = {Multi‐site feasibility and reproducibility study on UTE 3D phosphorous MRSI using novel rosette trajectory ( PETALUTE )},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30640}
}