Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 03.07.2025. Verfasst von Thomas H M Roos und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Oncology, Center for Image Sciences, University Medical Center Utrecht, Utrecht, the Netherlands. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Health˜Holland, Topsector Life Sciences and Health, University Medical Center Utrecht Brain Center, Young Talent Fellowship 2019, Toegepaste en Technische Wetenschappen, NWO, Talent Programme Vidi TTW 2019.
Abstract (Originalsprache)
Purpose
To enhance the accessibility of advanced pulse sequences, or parts thereof, through the open‐source Pulseq framework. This work extends the Pulseq framework to Philips MRI systems and incorporates dynamic parallel‐transmit (pTx) capabilities within the constraints of the existing Pulseq format. This enables the hybrid use of Pulseq sequences within the native vendor's scans, leveraging the combined strengths of both approaches. We showcase a new possibility of these techniques: the use of a portable cross‐vendor universal pulse (UP) excitation in a native scan. Methods pTx‐Pulseq was implemented to add full dynamic pTx within the Pulseq specification. We developed a Pulseq interpreter for Philips systems, supporting both Pulseq‐only and hybrid sequences. The hybrid mode was used to integrate a UP from the PASTeUR package into a vendor's native scan. Simulation experiments, safety validations, field measurements, and imaging experiments in phantom and in vivo were conducted to verify the interpreter's and UP's functionality.
Results
The interpreter executed Pulseq sequences on a Philips 7T system, accurately reproducing gradient waveforms and dynamic pTx sequences. The real‐time safety systems operated correctly. Phantom and in vivo scans demonstrated comparable image quality to native sequences, validating the effectiveness of the interpreter and the successful cross‐vendor use of universal pulses.
Conclusion
The successful cross‐vendor application of a universal pulse through pTx‐Pulseq in a hybrid sequence demonstrates how advanced MRI techniques can be made accessible. This not only highlights the flexibility and extensibility of Pulseq but also sets the stage for rapid clinical translation of innovative imaging techniques.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology and Oncology, Center for Image Sciences, University Medical Center Utrecht, Utrecht, the Netherlands.
- University of Paris-Saclay, CEA, CNRS, NeuroSpin, BAOBAB, Gif sur Yvette, France.
- Spinoza Center for Neuroimaging Amsterdam, Amsterdam, the Netherlands.
ORCID-Profile: Thomas H M Roos, Edwin Versteeg, Mark Gosselink, Hans Hoogduin, Kyung Min Nam, Nicolas Boulant, Vincent Gras, Franck Mauconduit, Dennis W J Klomp, Jeroen C W Siero, Jannie P Wijnen
Förderung
- Health˜Holland, Topsector Life Sciences and Health
- University Medical Center Utrecht Brain Center, Young Talent Fellowship 2019
- Toegepaste en Technische Wetenschappen, NWO, Talent Programme Vidi TTW 2019
- HORIZON EUROPE European Innovation Council, 2020 Research and Innovation programme
- HORIZON EUROPE European Innovation Council
Zitieren
APA
Roos, T. H. M., Versteeg, E., Gosselink, M., Hoogduin, H., Nam, K. M., Boulant, N., Gras, V., Mauconduit, F., Klomp, D. W. J., Siero, J. C. W., & Wijnen, J. P. (2025). pTx‐Pulseq in hybrid sequences: Accessible and advanced hybrid open‐source MRI sequences on Philips scanners. Magnetic Resonance in Medicine, 94(5), 1946-1962. https://doi.org/10.1002/mrm.30601
Vancouver
Roos THM, Versteeg E, Gosselink M, Hoogduin H, Nam KM, Boulant N, et al. pTx‐Pulseq in hybrid sequences: Accessible and advanced hybrid open‐source MRI sequences on Philips scanners. Magnetic Resonance in Medicine. 2025;94(5):1946-1962. doi:10.1002/mrm.30601
BibTeX
@article{Roos2025,
author = {Roos, Thomas H M and Versteeg, Edwin and Gosselink, Mark and Hoogduin, Hans and Nam, Kyung Min and Boulant, Nicolas and Gras, Vincent and Mauconduit, Franck and Klomp, Dennis W J and Siero, Jeroen C W and Wijnen, Jannie P},
title = {pTx‐Pulseq in hybrid sequences: Accessible and advanced hybrid open‐source MRI sequences on Philips scanners},
journal = {Magnetic Resonance in Medicine},
year = {2025},
volume = {94},
number = {5},
pages = {1946--1962},
doi = {10.1002/mrm.30601}
}
EUVIMED