Multi‐Contrast Human Brain CEST MRI at 11.7 T: First In Vivo Demonstration
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70498 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Camélia Ressam und 11 weiteren Autor:innen. Beteiligte Einrichtung: Université Paris-Saclay, CEA, CNRS, NeuroSpin, UMR 9027 BAOBAB, Gif-sur-Yvette, France. Hauptthemen laut MEDLINE: Brain, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 62 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Agence Nationale de la Recherche, HORIZON EUROPE European Innovation Council.
Abstract (Originalsprache)
Purpose
To implement and evaluate chemical exchange saturation transfer (CEST) at ultra‐high field on a whole‐body 11.7 T scanner using parallel transmission (pTx).
Methods
Tailored pTx CEST saturation pulses were optimized to achieve a spatially uniform target B 1 + while satisfying SAR and power constraints either defined by conservatively defined radio‐frequency (RF) power limits or by a virtual observation point (VOP)–based framework. The resulting pulse performance at 11.7 T was assessed through numerical simulations and compared against corresponding designs at 7 T. The proposed saturation pulses were integrated into a two‐dimensional multi‐slice EPI sequence. Experimental testing of the sequence was performed in vitro using metabolite‐doped phantoms and in vivo in four healthy volunteers.
Results
The pTx pulse designs optimized under VOP‐based SAR constraints at 11.7 T achieved comparable performance to 7 T designs despite higher intrinsic B 1 + inhomogeneity. In vitro experiments validated the sequence implementation and confirmed detection of the targeted metabolites. In vivo experiments demonstrated acceptable saturation homogeneity (nRMSE ≤ 12.1%) in a slab of thickness 1 cm, with only minimal additional improvement achieved through B 1 + post‐processing correction. Across the cohort, robust gray matter–white matter contrast was consistently observed for multiple CEST contrasts.
Conclusion
VOP‐based SAR‐constrained pTx CEST saturation pulse design enabled robust and reproducible contrast at 11.7 T by effectively mitigating B 1 + inhomogeneities while remaining within safety limits, facilitating translation to future in vivo clinical research applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Université Paris-Saclay, CEA, CNRS, NeuroSpin, UMR 9027 BAOBAB, Gif-sur-Yvette, France.
- Siemens Healthcare SAS, Courbevoie, France.
- Imaging Center of Excellence, University of Glasgow, Glasgow, UK.
ORCID-Profile: Camélia Ressam, Alexis Amadon, Alexandre Vignaud, Son Chu, Shajan Gunamony, Nicolas Boulant, Vincent Gras, Luisa Ciobanu
Förderung
- Agence Nationale de la Recherche
- HORIZON EUROPE European Innovation Council
Zitieren
APA
Ressam, C., Brégeat, J., Lopez Kolkovsky, A. L., Amadon, A., Vignaud, A., Massire, A., Chu, S., Gunamony, S., Boulant, N., Gras, V., Mauconduit, F., & Ciobanu, L. (2026). Multi‐Contrast Human Brain CEST MRI at 11.7 T: First In Vivo Demonstration. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70498
Vancouver
Ressam C, Brégeat J, Lopez Kolkovsky AL, Amadon A, Vignaud A, Massire A, et al. Multi‐Contrast Human Brain CEST MRI at 11.7 T: First In Vivo Demonstration. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70498
BibTeX
@article{Ressam2026,
author = {Ressam, Camélia and Brégeat, Joseph and Lopez Kolkovsky, Alfredo L and Amadon, Alexis and Vignaud, Alexandre and Massire, Aurélien and Chu, Son and Gunamony, Shajan and Boulant, Nicolas and Gras, Vincent and Mauconduit, Franck and Ciobanu, Luisa},
title = {Multi‐Contrast Human Brain CEST MRI at 11.7 T: First In Vivo Demonstration},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70498}
}
EUVIMED