Whole liver phase‐based R2 mapping in liver iron overload within a breath‐hold
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30461 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2025. Written by Daiki Tamada and 4 co-authors. Institution: Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA. Major MEDLINE topics: Iron Overload, Liver, Breath Holding, Magnetic Resonance Imaging. Cited 2 times so far (Crossref); 44 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIDDK NIH HHS, NCATS NIH HHS, NIBIB NIH HHS.
Abstract (original language)
Purpose
Diagnosis and treatment monitoring of iron overload increasingly relies on non‐invasive MRI‐based measurement of liver iron concentration (LIC). Liver R 2 mapping is known to correlate well with LIC. However, traditional spin‐echo based R 2 mapping methods have drawbacks such as long acquisition times and limited volumetric coverage. In this work, we present an optimized phase‐based R 2 mapping method to quantify whole‐liver R 2 in iron overload patients within a single breathhold. Theory and Methods A recently developed phase‐based R 2 mapping method is optimized in this study to improve estimation of high R 2 values using reduced TR, spatial averaging, and R 1 correction. Using Bloch equation simulations, the proposed optimization method was evaluated. Furthermore, the impact of fat on R 2 bias was investigated through simulations. The feasibility of the optimized phase‐based R 2 method was assessed in healthy volunteers and patients with iron overload and compared to reference STEAM‐MRS R 2 measurements.
Results
Simulations demonstrate that a shorter TR extends the dynamic range of R 2 estimation to higher values and that averaging of signal phase before R 2 estimation is necessary when R 2 is high. Phantom experiments also demonstrate reduced phase‐based R 2 bias using R 1 correction. Good agreement (1.5 T: r 2 = 0.76, 3.0 T: r 2 = 0.70) between the modified phase‐based R 2 method and reference STEAM R 2 was found in healthy volunteers and iron overload patients over a wide range of LIC values.
Conclusion
This study demonstrates the feasibility of the proposed phase‐based R 2 method to accurately measure whole‐liver R 2 mapping in severe iron overloaded patients during a single breathhold.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Medicine, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Emergency Medicine, University of Wisconsin-Madison, Madison, Wisconsin, USA.
ORCID profiles: Daiki Tamada, Diego Hernando, Scott B Reeder
Funding
- NIDDK NIH HHS
- NCATS NIH HHS
- NIBIB NIH HHS
- GE Healthcare
- National Institute of Diabetes and Digestive and Kidney Diseases
- National Center for Advancing Translational Sciences
- National Institute of Biomedical Imaging and Bioengineering
Cite
APA
Tamada, D., Navaratna, R., Weaver, J. M., Hernando, D., & Reeder, S. B. (2025). Whole liver phase‐based R2 mapping in liver iron overload within a breath‐hold. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30461
Vancouver
Tamada D, Navaratna R, Weaver JM, Hernando D, Reeder SB. Whole liver phase‐based R2 mapping in liver iron overload within a breath‐hold. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30461
BibTeX
@article{Tamada2025,
author = {Tamada, Daiki and Navaratna, Ruvini and Weaver, Jayse Merle and Hernando, Diego and Reeder, Scott B},
title = {Whole liver phase‐based R2 mapping in liver iron overload within a breath‐hold},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30461}
}
EUVIMED