Simultaneous T1, T2 and T2⁎ mapping of the liver with multi-shot MI-SAGE
Magnetic Resonance Imaging · 2024
DOI: 10.1016/j.mri.2023.11.004 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2024. Verfasst von Ke Zhang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic & Interventional Radiology, Heidelberg University Hospital, Heidelberg, Germany; Translational Lung Research Center (TLRC), …. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Liver. Die Arbeit nennt 25 Literaturangaben. Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
PURPOSE: To apply multi-shot high-resolution multi inversion spin and gradient echo (MI-SAGE) acquisition for simultaneous liver T1, T2 and T2* mapping. METHODS: Inversion prepared spin- and gradient-echo EPI was developed with ascending slice order across measurements for efficient acquisition with T1, T2, and T2⁎ weighting. Multi-shot EPI was also implemented to minimize distortion and blurring while enabling high in-plane resolution. A dictionary-matching approach was used to fit the images to quantitative parameter maps, which were compared to T1 measured by modified Look-Locker (MOLLI), T1 measured by variable flip angle (VFA), T2 measured by multiple echo time-based Half Fourier Single-shot Turbo spin-Echo (HASTE), T2 measured by radial turbo-spin-echo (rTSE) and T2⁎ measured by multiple gradient echo (MGRE) sequences. RESULTS: The multi-shot variant of the sequence achieved higher in-plane resolution of 1.7 × 1.7 mm2 with good image quality in 28 s. Derived quantitative maps showed comparable values to conventional mapping methods. As measured in phantom and in vivo, MOLLI, MESE and MGRE give closest values to MISAGE. VFA, HASTE and rTSE show obvious overestimation. CONCLUSIONS: The proposed multi-shot inversion prepared spin- and gradient-echo EPI sequence allows for high-resolution quantitative T1, T2 and T2 liver tissue characterization in a single breath-hold scan.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Diagnostic & Interventional Radiology, Heidelberg University Hospital, Heidelberg, Germany; Translational Lung Research Center (TLRC), German Center for Lung Research (DZL), Heidelberg, Germany; Department…
- Divison of Radiology, German Cancer Research Center, Heidelberg, Germany.
- Divison of Medical Physics in Radiology, German Cancer Research Center, Heidelberg, Germany; Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany; Faculty of Medicine, Heidelberg University, Heid…
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Zhang, K., Triphan, S. M. F., Wielpütz, M. O., Ziener, C. H., Ladd, M. E., Schlemmer, H. P., Kauczor, H. U., Kurz, F. T., & Sedlaczek, O. (2024). Simultaneous T1, T2 and T2⁎ mapping of the liver with multi-shot MI-SAGE. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2023.11.004
Vancouver
Zhang K, Triphan SMF, Wielpütz MO, Ziener CH, Ladd ME, Schlemmer HP, et al. Simultaneous T1, T2 and T2⁎ mapping of the liver with multi-shot MI-SAGE. Magnetic Resonance Imaging. 2024. doi:10.1016/j.mri.2023.11.004
BibTeX
@article{Zhang2024,
author = {Zhang, Ke and Triphan, Simon M F and Wielpütz, Mark O and Ziener, Christian H and Ladd, Mark E and Schlemmer, Heinz-Peter and Kauczor, Hans-Ulrich and Kurz, Felix T and Sedlaczek, Oliver},
title = {Simultaneous T1, T2 and T2⁎ mapping of the liver with multi-shot MI-SAGE},
journal = {Magnetic Resonance Imaging},
year = {2024},
doi = {10.1016/j.mri.2023.11.004}
}
EUVIMED