Mechanically anisotropic phantoms for magnetic resonance elastography
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30394 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Kevin N Eckstein und 8 weiteren Autor:innen. Beteiligte Einrichtung: Mechanical Engineering and Materials Science, Washington University, St. Louis, Missouri, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Elasticity Imaging Techniques, Magnetic Resonance Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 71 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIH HHS, NIBIB NIH HHS, Office of Naval Research.
Abstract (Originalsprache)
Purpose
Imaging phantoms with known anisotropic mechanical properties are needed to evaluate magnetic resonance elastography (MRE) methods to estimate anisotropic parameters. The aims of this study were to fabricate mechanically anisotropic MRE phantoms, characterize their mechanical behavior by direct testing, then assess the accuracy of MRE estimates of anisotropic properties using a transversely isotropic nonlinear inversion (TI‐NLI) algorithm.
Methods
Directionally scaled and unscaled lattices were designed to exhibit anisotropic or isotropic mechanical properties. Lattices were three‐dimensionally printed in poly(ethelyne glycol) diacrylate using a commercial digital light processing printer, then infilled with gelatin to form a composite material. Benchtop testing determined two shear stiffnesses, and , governing loading parallel and perpendicular to the symmetry axis, and two analogous Young's moduli and . From these measures, shear anisotropy = and tensile anisotropy = were calculated. Three phantoms were driven by a central actuator and imaged with MRE at frequencies from 300 to 500 Hz. From MRE data, the TI‐NLI algorithm estimated maps of , , and .
Results
In benchtop tests, geometrically scaled lattice composites exhibited the following anisotropic properties: = 6.1 ± 0.7 kPa, = 0.83 ± 0.13, = 0.78 ± 0.09} (mean ± standard deviation). MRE of scaled lattice composites revealed elliptical wavefields; TI‐NLI analysis identified the following median property ranges: = 11–19 kPa, = 0.6–1.0, = 0.8–1.6}.
Conclusion
Anisotropic MRE phantoms are created by embedding anisotropic three‐dimensionally printed lattices into a softer matrix. The TI‐NLI algorithm accurately estimates spatial contrast in anisotropic properties.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Mechanical Engineering and Materials Science, Washington University, St. Louis, Missouri, USA.
- Biomedical Engineering, University of Delaware, Newark, Delaware, USA.
- Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA.
- Biomedical Engineering, Washington University, St. Louis, Missouri, USA.
ORCID-Profile: Kevin N Eckstein, Daniel Yoon, Curtis L Johnson, Philip V Bayly
Förderung
- NIH HHS
- NIBIB NIH HHS
- Office of Naval Research
- National Institutes of Health
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Eckstein, K. N., Yoon, D., Ruding, M., Balouchzadeh, R., Thompson-Mazzeo, A., Okamoto, R. J., Johnson, C. L., McGarry, M. D. J., & Bayly, P. V. (2025). Mechanically anisotropic phantoms for magnetic resonance elastography. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30394
Vancouver
Eckstein KN, Yoon D, Ruding M, Balouchzadeh R, Thompson-Mazzeo A, Okamoto RJ, et al. Mechanically anisotropic phantoms for magnetic resonance elastography. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30394
BibTeX
@article{Eckstein2025,
author = {Eckstein, Kevin N and Yoon, Daniel and Ruding, Margrethe and Balouchzadeh, Ramin and Thompson-Mazzeo, Aaliyah and Okamoto, Ruth J and Johnson, Curtis L and McGarry, Matthew D J and Bayly, Philip V},
title = {Mechanically anisotropic phantoms for magnetic resonance elastography},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30394}
}
EUVIMED