EUVIMEDEUVIMED entdecken
Crossref

ORACLE: An analytical approach for T 1 , T 2 , proton density, and off‐resonance mapping with phase‐cycled balanced steady‐state free precession

Nils M J Plähn iD, Yasaman Safarkhanlo iD, Berk C Açikgöz, Adèle L C Mackowiak, Piotr Radojewski, Gabriele Bonanno, Eva S Peper, Rahel Heule, Jessica A M Bastiaansen iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30388 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Nils M J Plähn und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic, Interventional and Pediatric Radiology (DIPR), Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 57 Literaturangaben. Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop and validate a novel analytical approach simplifying , , proton density (PD), and off‐resonance quantifications from phase‐cycled balanced steady‐state free precession (bSSFP) data. Additionally, to introduce a method to correct aliasing effects in undersampled bSSFP profiles. Theory and Methods Off‐resonant‐encoded analytical parameter quantification using complex linearized equations (ORACLE) provides analytical solutions for bSSFP profiles. which instantaneously quantify , , proton density (PD), and . An aliasing correction formalism was derived to allow undersampling of bSSFP profiles. ORACLE was used to quantify , , PD, /, and based on fully sampled () bSSFP profiles from numerical simulations and 3T MRI experiments in phantom and 10 healthy subjects' brains. Obtained values were compared with reference scans in the same scan session. Aliasing correction was validated in subsampled () bSSFP profiles in numerical simulations and human brains.

Results

ORACLE quantifications agreed well with input values from simulations and phantom reference values ( R 2 = 0.99). In human brains, and quantifications when compared with reference methods showed coefficients of variation below 2.9% and 3.9%, biases of 182 and 16.6 ms, and mean white‐matter values of 642 and 51 ms using ORACLE. The quantification differed less than 3 Hz between both methods. PD and maps had comparable histograms. The maps effectively identified cerebrospinal fluid. Aliasing correction removed aliasing‐related quantification errors in undersampled bSSFP profiles, significantly reducing scan time.

Conclusion

ORACLE enables simplified and rapid quantification of , , PD, and from phase‐cycled bSSFP profiles, reducing acquisition time and eliminating biomarker maps' coregistration issues.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingBrainImage Processing, Computer-AssistedHumansAlgorithmsPhantoms, ImagingProtonsReproducibility of ResultsComputer SimulationAdult

Stichwörter der Autor:innen

aliasing correctionanalytical solutionsbrainphase‐cycled bSSFPquantitative MRI

Substanzen

Protons

Autor:innen und Institutionen

  • Department of Diagnostic, Interventional and Pediatric Radiology (DIPR), Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Translational Imaging Center (TIC), Swiss Institute for Translational and Entrepreneurial Medicine, Bern, Switzerland.
  • Graduate School for Cellular and Biomedical Sciences (GCB), University of Bern, Bern, Switzerland.
  • Department of Cardiology, Inselspital, University Hospital Bern, Bern, Switzerland.
  • Department of Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
  • University Institute of Diagnostic and Interventional Neuroradiology, Bern, Switzerland.

ORCID-Profile: Nils M J Plähn, Yasaman Safarkhanlo, Jessica A M Bastiaansen

Förderung

  • Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung

Zitieren

APA

Plähn, N. M. J., Safarkhanlo, Y., Açikgöz, B. C., Mackowiak, A. L. C., Radojewski, P., Bonanno, G., Peper, E. S., Heule, R., & Bastiaansen, J. A. M. (2025). ORACLE: An analytical approach for T 1 , T 2 , proton density, and off‐resonance mapping with phase‐cycled balanced steady‐state free precession. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30388

Vancouver

Plähn NMJ, Safarkhanlo Y, Açikgöz BC, Mackowiak ALC, Radojewski P, Bonanno G, et al. ORACLE: An analytical approach for T 1 , T 2 , proton density, and off‐resonance mapping with phase‐cycled balanced steady‐state free precession. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30388

BibTeX
@article{Plhn2025,
  author = {Plähn, Nils M J and Safarkhanlo, Yasaman and Açikgöz, Berk C and Mackowiak, Adèle L C and Radojewski, Piotr and Bonanno, Gabriele and Peper, Eva S and Heule, Rahel and Bastiaansen, Jessica A M},
  title = {ORACLE: An analytical approach for T 1 , T 2 , proton density, and off‐resonance mapping with phase‐cycled balanced steady‐state free precession},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},