Custom integration of a magnetic‐field monitoring system into a 32‐channel MRI head coil

Tim Schmidt iD, Yoojin Lee, Zoltan Nagy iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30314 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Tim Schmidt und 2 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory for Social and Neural Systems Research (SNS Lab), University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Equipment Design, Brain. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NCRR NIH HHS, NINDS NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Customizing a Siemens 32‐channel coil for use in a Philips 3T MRI system with incorporated magnetic field probes for collecting high‐quality MRI and magnetic‐field monitoring data concurrently.

Methods

The development process of the custom coil involved several (iterative) phases. Standard temporal SNR and B 1 + data were collected with the 32‐channel Siemens and for reference the 32‐channel/8‐channel Philips head coils before and after the custom coil was made compatible with the 3T Philips Achieva system, and magnetic field probes were installed into it along with ancillary electronics around it. Quality assurance tests were conducted in each of the build phases to ensure that the modifications did not affect MRI or field‐monitoring data negatively. To test the finished custom coil, we collected high angular resolution diffusion imaging (HARDI) datasets on a spherical silicon oil phantom both with and without concurrent field monitoring and a 32‐channel Philips coil without concurrent field monitoring, where the latter two served as reference scans to assess the improved performance of the custom coil with field monitoring. Similar HARDI‐MRI data were also collected in vivo with the finished custom coil together with field monitoring data.

Results

The custom coil provided excellent temporal SNR especially at the edges where cortical gray matter is expected. When using concurrent field monitoring in HARDI acquisitions, the custom coil alleviated ghosting artifacts in phantom data and provided in vivo images with 1.4‐mm isotropic resolution.

Conclusion

The custom MRI coil with integrated magnetic‐field monitoring probes provided improved imaging performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingEquipment DesignBrainHeadSignal-To-Noise RatioMagnetic FieldsHumansImage Processing, Computer-AssistedMaleAdultFemale

Stichwörter der Autor:innen

MRIhead coilindustrial designmagnetic field monitoring

Autor:innen und Institutionen

  • Laboratory for Social and Neural Systems Research (SNS Lab), University of Zurich, Zurich, Switzerland.
  • Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.

ORCID-Profile: Tim Schmidt, Zoltan Nagy

Förderung

  • NIBIB NIH HHS
  • NCRR NIH HHS
  • NINDS NIH HHS

Zitieren

APA

Schmidt, T., Lee, Y., & Nagy, Z. (2025). Custom integration of a magnetic‐field monitoring system into a 32‐channel MRI head coil. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30314

Vancouver

Schmidt T, Lee Y, Nagy Z. Custom integration of a magnetic‐field monitoring system into a 32‐channel MRI head coil. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30314

BibTeX
@article{Schmidt2025,
  author = {Schmidt, Tim and Lee, Yoojin and Nagy, Zoltan},
  title = {Custom integration of a magnetic‐field monitoring system into a 32‐channel MRI head coil},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30314}
}
Kostenlos starten →