T1 and T2 measurements across multiple 0.55T MRI systems using open‐source vendor‐neutral sequences

Kathryn E Keenan iD, Bilal Tasdelen iD, Ahsan Javed iD, Rajiv Ramasawmy, Rudy Rizzo, Michele N Martin, Karl F Stupic, Nicole Seiberlich, Adrienne E Campbell-Washburn iD, Krishna S Nayak iD

Magnetic Resonance in Medicine · 2025-01 · 93 (1) · 289-300

DOI: 10.1002/mrm.30281 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024-09. Verfasst von Kathryn E Keenan und 9 weiteren Autor:innen. Beteiligte Einrichtung: National Institute of Standards and Technology, Boulder, Colorado, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Intramural NIH HHS, National Institutes of Health, Intramural Research Program, National Heart Lung and Blood Institute, National Science Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare T1 and T2 measurements across commercial and prototype 0.55T MRI systems in both phantom and healthy participants using the same vendor‐neutral pulse sequences, reconstruction, and analysis methods.

Methods

Standard spin echo measurements and abbreviated protocol measurements of T1, B1, and T2 were made on two prototype 0.55 T systems and two commercial 0.55T systems using an ISMRM/NIST system phantom. Additionally, five healthy participants were imaged at each system using the abbreviated protocol for T1, B1, and T2 measurement. The phantom measurements were compared to NMR‐based reference measurements to determine accuracy, and both phantom and in vivo measurements were compared to assess reproducibility and differences between the prototype and commercial systems.

Results

Vendor‐neutral sequences were implemented across all four systems, and the code for pulse sequences and reconstruction is freely available. For participants, there was no difference in the mean T1 and T2 relaxation times between the prototype and commercial systems. In the phantom, there were no significant differences between the prototype and commercial systems for T1 and T2 measurements using the abbreviated protocol.

Conclusion

Quantitative T1 and T2 measurements at 0.55T in phantom and healthy participants are not statistically different across the prototype and commercial systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingPhantoms, ImagingHumansReproducibility of ResultsMaleAdultFemaleAlgorithmsImage Interpretation, Computer-AssistedImage Processing, Computer-AssistedEquipment DesignSensitivity and Specificity

Stichwörter der Autor:innen

0.55TT1T2multi‐site repeatability

Autor:innen und Institutionen

  • National Institute of Standards and Technology, Boulder, Colorado, USA.
  • Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
  • Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, Blood Institute, National Institutes of Health, Bethesda, Maryland, USA.
  • Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.

ORCID-Profile: Kathryn E Keenan, Bilal Tasdelen, Ahsan Javed, Adrienne E Campbell-Washburn, Krishna S Nayak

Förderung

  • Intramural NIH HHS
  • National Institutes of Health, Intramural Research Program, National Heart Lung and Blood Institute
  • National Science Foundation

Zitieren

APA

Keenan, K. E., Tasdelen, B., Javed, A., Ramasawmy, R., Rizzo, R., Martin, M. N., Stupic, K. F., Seiberlich, N., Campbell-Washburn, A. E., & Nayak, K. S. (2025). T1 and T2 measurements across multiple 0.55T MRI systems using open‐source vendor‐neutral sequences. Magnetic Resonance in Medicine, 93(1), 289-300. https://doi.org/10.1002/mrm.30281

Vancouver

Keenan KE, Tasdelen B, Javed A, Ramasawmy R, Rizzo R, Martin MN, et al. T1 and T2 measurements across multiple 0.55T MRI systems using open‐source vendor‐neutral sequences. Magnetic Resonance in Medicine. 2025;93(1):289-300. doi:10.1002/mrm.30281

BibTeX
@article{Keenan2025,
  author = {Keenan, Kathryn E and Tasdelen, Bilal and Javed, Ahsan and Ramasawmy, Rajiv and Rizzo, Rudy and Martin, Michele N and Stupic, Karl F and Seiberlich, Nicole and Campbell-Washburn, Adrienne E and Nayak, Krishna S},
  title = {T1 and T2 measurements across multiple 0.55T MRI systems using open‐source vendor‐neutral sequences},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  volume = {93},
  number = {1},
  pages = {289--300},
  doi = {10.1002/mrm.30281}
}
Kostenlos starten →