Accelerated 3D multi‐echo spin‐echo sequence with a subspace constrained reconstruction for whole mouse brain T2 mapping
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30146 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Aurélien J Trotier und 3 weiteren Autor:innen. Beteiligte Einrichtung: Centre de Résonance Magnétique des Systèmes Biologiques, UMR5536, CNRS, University Bordeaux, Bordeaux, France. Hauptthemen laut MEDLINE: Brain, Imaging, Three-Dimensional, Phantoms, Imaging, Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: Ligue Contre le Cancer, French Government in the framework of the University of Bordeaux's France 2030 program/RRI "IMPACT".
Abstract (Originalsprache)
Purpose
To accelerate whole‐brain quantitative mapping in preclinical imaging setting.
Methods
A three‐dimensional (3D) multi‐echo spin echo sequence was highly undersampled with a variable density Poisson distribution to reduce the acquisition time. Advanced iterative reconstruction based on linear subspace constraints was employed to recover high‐quality raw images. Different subspaces, generated using exponential or extended‐phase graph (EPG) simulations or from low‐resolution calibration images, were compared. The subspace dimension was investigated in terms of precision. The method was validated on a phantom containing a wide range of and was then applied to monitor metastasis growth in the mouse brain at 4.7T. Image quality and estimation were assessed for 3 acceleration factors (6/8/10).
Results
The EPG‐based dictionary gave robust estimations of a large range of . A subspace dimension of 6 was the best compromise between precision and image quality. Combining the subspace constrained reconstruction with a highly undersampled dataset enabled the acquisition of whole‐brain maps, the detection and the monitoring of metastasis growth of less than 500 .
Conclusion
Subspace‐based reconstruction is suitable for 3D mapping. This method can be used to reach an acceleration factor up to 8, corresponding to an acquisition time of 25 min for an isotropic 3D acquisition of 156 m on the mouse brain, used here for monitoring metastases growth.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre de Résonance Magnétique des Systèmes Biologiques, UMR5536, CNRS, University Bordeaux, Bordeaux, France.
ORCID-Profile: Aurélien J Trotier, Nadège Corbin, Sylvain Miraux, Emeline J Ribot
Förderung
- Ligue Contre le Cancer
- French Government in the framework of the University of Bordeaux's France 2030 program/RRI "IMPACT"
Zitieren
APA
Trotier, A. J., Corbin, N., Miraux, S., & Ribot, E. J. (2024). Accelerated 3D multi‐echo spin‐echo sequence with a subspace constrained reconstruction for whole mouse brain T2 mapping. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30146
Vancouver
Trotier AJ, Corbin N, Miraux S, Ribot EJ. Accelerated 3D multi‐echo spin‐echo sequence with a subspace constrained reconstruction for whole mouse brain T2 mapping. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30146