EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Fast and High‐Resolution T 2 Mapping Based on Echo Merging Plus k ‐t Undersampling with Reduced Refocusing Flip Angles ( TEMPURA ) as Methods for Human Renal MRI

Hao Li iD, Andrew N Priest iD, Ines Horvat-Menih, Yuan Huang, Shaohang Li, Grant D Stewart, Iosif A Mendichovszky, Susan T Francis, Ferdia A Gallagher

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30115 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Hao Li und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Science and Technology for Brain-Inspired Intelligence, Fudan University, Shanghai, China. Hauptthemen laut MEDLINE: Algorithms, Kidney Neoplasms, Phantoms, Imaging, Kidney. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Shanghai Municipal Science and Technology Major Project, Shanghai Center for Brain Science and Brain-Inspired Technology, The Mark Foundation for Cancer Research, the CRUK Cambridge Centre.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop a highly accelerated multi‐echo spin‐echo method, TEMPURA, for reducing the acquisition time and/or increasing spatial resolution for kidney T 2 mapping.

Methods

TEMPURA merges several adjacent echoes into one k ‐space by either combining independent echoes or sharing one echo between k ‐spaces. The combined k ‐space is reconstructed based on compressed sensing theory. Reduced flip angles are used for the refocusing pulses, and the extended phase graph algorithm is used to correct the effects of indirect echoes. Two sequences were developed: a fast breath‐hold sequence; and a high‐resolution sequence. The performance was evaluated prospectively on a phantom, 16 healthy subjects, and two patients with different types of renal tumors.

Results

The fast TEMPURA method reduced the acquisition time from 3–5 min to one breath‐hold (18 s). Phantom measurements showed that fast TEMPURA had a mean absolute percentage error (MAPE) of 8.2%, which was comparable to a standardized respiratory‐triggered sequence (7.4%), but much lower than a sequence accelerated by purely k ‐t undersampling (21.8%). High‐resolution TEMPURA reduced the in‐plane voxel size from 3 × 3 to 1 × 1 mm 2 , resulting in improved visualization of the detailed anatomical structure. In vivo T 2 measurements demonstrated good agreement (fast: MAPE = 1.3%–2.5%; high‐resolution: MAPE = 2.8%–3.3%) and high correlation coefficients (fast: R = 0.85–0.98; high‐resolution: 0.82–0.96) with the standardized method, outperforming k ‐t undersampling alone (MAPE = 3.3–4.5%, R = 0.57–0.59).

Conclusion

TEMPURA provides fast and high‐resolution renal T 2 measurements. It has the potential to improve clinical throughput and delineate intratumoral heterogeneity and tissue habitats at unprecedented spatial resolution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsKidney NeoplasmsPhantoms, ImagingKidneyHumansMagnetic Resonance ImagingFemaleAdultMaleImage Interpretation, Computer-AssistedReproducibility of ResultsMiddle AgedImage EnhancementImage Processing, Computer-AssistedBreath Holding

Stichwörter der Autor:innen

T2 mappingcompressed sensingimaging accelerationkidneymulti‐echo spin‐echoquantitative MRI

Autor:innen und Institutionen

  • Institute of Science and Technology for Brain-Inspired Intelligence, Fudan University, Shanghai, China.
  • Department of Radiology, University of Cambridge, Cambridge, UK.
  • Department of Radiology, Cambridge University Hospitals NHS Foundation Trust, Addenbrooke's Hospital, Cambridge, UK.
  • EPSRC Cambridge Mathematics of Information in Healthcare Hub, University of Cambridge, Cambridge, UK.
  • Department of Surgery, Cambridge University Hospitals NHS Foundation Trust, Addenbrooke's Hospital, Cambridge, UK.
  • Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, UK.

ORCID-Profile: Hao Li, Andrew N Priest

Förderung

  • Shanghai Municipal Science and Technology Major Project
  • Shanghai Center for Brain Science and Brain-Inspired Technology
  • The Mark Foundation for Cancer Research, the CRUK Cambridge Centre
  • Shanghai Pujiang Program
  • National Natural Science Foundation of China
  • Addenbrooke's Charitable Trust, Cambridge University Hospitals
  • Medical Research Council
  • CRUK National Cancer Imaging Translational Accelerator award (NCITA)

Zitieren

APA

Li, H., Priest, A. N., Horvat-Menih, I., Huang, Y., Li, S., Stewart, G. D., Mendichovszky, I. A., Francis, S. T., & Gallagher, F. A. (2024). Fast and High‐Resolution T 2 Mapping Based on Echo Merging Plus k ‐t Undersampling with Reduced Refocusing Flip Angles ( TEMPURA ) as Methods for Human Renal MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30115

Vancouver

Li H, Priest AN, Horvat-Menih I, Huang Y, Li S, Stewart GD, et al. Fast and High‐Resolution T 2 Mapping Based on Echo Merging Plus k ‐t Undersampling with Reduced Refocusing Flip Angles ( TEMPURA ) as Methods for Human Renal MRI. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30115

BibTeX
@article{Li2024,
  author = {Li, Hao and Priest, Andrew N and Horvat-Menih, Ines and Huang, Yuan and Li, Shaohang and Stewart, Grant D and Mendichovszky, Iosif A and Francis, Susan T and Gallagher, Ferdia A},
  title = {Fast and High‐Resolution T 2 Mapping Based on Echo Merging Plus k ‐t Undersampling with Reduced Refocusing Flip Angles ( TEMPURA ) as Methods for Human Renal MRI},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30115}
}
Kostenlos starten →