Improvements in Between‐Vendor MRI Harmonization of Renal T 2 Mapping using Stimulated Echo Compensation

Hao Li iD, Alexander J Daniel iD, Charlotte E Buchanan iD, Fábio Nery, David M Morris, Shaohang Li iD, Yuan Huang, João A Sousa iD, Steven Sourbron iD, Iosif A Mendichovszky iD, David L Thomas iD, Andrew N Priest iD und 1 weitere

Journal of Magnetic Resonance Imaging · 2024

DOI: 10.1002/jmri.29282 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2024. Verfasst von Hao Li und 12 weiteren Autor:innen. Beteiligte Einrichtung: The Institute of Science and Technology for Brain-inspired Intelligence, Fudan University, Shanghai, China. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging, Kidney, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIHR Cambridge Biomedical Research Centre, Medical Research Council, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

T 2 mapping is valuable to evaluate pathophysiology in kidney disease. However, variations in T 2 relaxation time measurements across MR scanners and vendors may occur requiring additional correction.

Purpose

To harmonize renal T 2 measurements between MR vendor platforms, and use an extended‐phase‐graph‐based fitting method (“StimFit”) to correct stimulated echoes and reduce between‐vendor variations. Study Type Prospective. Subjects 8 healthy “travelling” volunteers (37.5% female, 32 ± 6 years) imaged on four MRI systems across three vendors at four sites, 10 healthy volunteers (50% female, 32 ± 8 years) scanned multiple times on a given MR scanner for repeatability evaluation. ISMRM/NIST system phantom scanned for evaluation of T 2 accuracy. Field Strength/Sequence 3T, multiecho spin‐echo sequence. Assessment T 2 images fit using conventional monoexponential fitting and “StimFit.” Mean absolute percentage error (MAPE) of phantom measurements with reference T 2 values. Average cortex and medulla T 2 values compared between MR vendors, with masks obtained from T 2 ‐weighted images and T 1 maps. Full‐width‐at‐half‐maximum (FWHM) T 2 distributions to evaluate local homogeneity of measurements. Statistical Tests Coefficient of variation (CV), linear mixed‐effects model, analysis of variance, student's t ‐tests, Bland–Altman plots, P ‐value

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingPhantoms, ImagingKidneyImage Processing, Computer-AssistedHumansFemaleAdultProspective StudiesMaleReproducibility of ResultsHealthy VolunteersAlgorithms

Stichwörter der Autor:innen

T2 mappingextended phase graphharmonizationkidneymultiecho spin‐echostimulated echo compensation

Autor:innen und Institutionen

  • The Institute of Science and Technology for Brain-inspired Intelligence, Fudan University, Shanghai, China.
  • Department of Radiology, University of Cambridge, Cambridge, UK.
  • Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, UK.
  • Developmental Imaging and Biophysics Section, UCL Great Ormond Street Institute of Child Health, London, UK.
  • Centre for Cardiovascular Science, University of Edinburgh, Edinburgh, UK.
  • EPSRC Cambridge Mathematics of Information in Healthcare Hub, University of Cambridge, Cambridge, UK.

ORCID-Profile: Hao Li, Alexander J Daniel, Charlotte E Buchanan, Shaohang Li, João A Sousa, Steven Sourbron, Iosif A Mendichovszky, David L Thomas, Andrew N Priest, Susan T Francis

Förderung

  • NIHR Cambridge Biomedical Research Centre
  • Medical Research Council
  • National Natural Science Foundation of China
  • Shanghai Pujiang Program
  • NIHR Nottingham Biomedical Research Centre
  • Medical Research Council Grant

Zitieren

APA

Li, H., Daniel, A. J., Buchanan, C. E., Nery, F., Morris, D. M., Li, S., Huang, Y., Sousa, J. A., Sourbron, S., Mendichovszky, I. A., Thomas, D. L., Priest, A. N., & Francis, S. T. (2024). Improvements in Between‐Vendor MRI Harmonization of Renal T 2 Mapping using Stimulated Echo Compensation. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29282

Vancouver

Li H, Daniel AJ, Buchanan CE, Nery F, Morris DM, Li S, et al. Improvements in Between‐Vendor MRI Harmonization of Renal T 2 Mapping using Stimulated Echo Compensation. Journal of Magnetic Resonance Imaging. 2024. doi:10.1002/jmri.29282

BibTeX
@article{Li2024,
  author = {Li, Hao and Daniel, Alexander J and Buchanan, Charlotte E and Nery, Fábio and Morris, David M and Li, Shaohang and Huang, Yuan and Sousa, João A and Sourbron, Steven and Mendichovszky, Iosif A and Thomas, David L and Priest, Andrew N and Francis, Susan T},
  title = {Improvements in Between‐Vendor MRI Harmonization of Renal T 2 Mapping using Stimulated Echo Compensation},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2024},
  doi = {10.1002/jmri.29282}
}
Kostenlos starten →