EUVIMEDEUVIMED entdecken
Open Access

Open‐source implementation of X‐nuclear sequences using the Pulseq framework

Xiaoxi Liu iD, Di Cui iD, Peder E Z Larson iD, Dirk Mayer, Andreas Korzowski, Jon-Fredrik Nielsen, Rolf F Schulte, Changhua Mu, Lucas Carvajal, Duan Xu, Jeremy W Gordon iD, Daniel B Vigneron und 2 weitere

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30509 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Xiaoxi Liu und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Biomedical Imaging, University of California, California, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIH HHS, UCSF Department of Radiology & Biomedical Imaging Seed Grant.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Create vendor‐neutral modular sequences for X‐nuclear acquisitions and build an X‐nuclear–enabled Pulseq interpreter for GE (GE HealthCare, Waukesha, WI) scanners.

Methods

We designed a modular 2D gradient echo spiral sequence to support several sequence formats and a modular metabolite‐specific 3D balanced steady‐state free precession sequence for hyperpolarized (HP) carbon‐13 ( 13 C) MRI. In addition, we developed a new Pulseq interpreter for GE scanners, named TOPPE MNS (TOPPE Multi‐Nuclear Spectroscopy), to implement X‐nuclear acquisitions capabilities. We evaluated TOPPE MNS and the modular sequences through phantom studies using phosphorus‐31 ( 31 P), hydrogen‐2 ( 2 H), and 13 C coils, and in vivo studies including a human brain deuterium metabolic imaging study at natural abundance, HP 13 C animal studies, and human renal studies.

Results

Data from the 13 C phantom showed the accuracy of designed modular sequences and consistent performance with the product sequences. 31 P, 2 H, and 13 C phantom studies and a multi‐vendor/multi‐version 13 C phantom study showed accurate excitation and spatial encoding functionalities. A 2 H‐MRS brain volunteer study, HP [1‐ 13 C]pyruvate animal study, and human renal study showed good image quality with SNR comparable to those reported in the published literature. These results demonstrated the reproducibility of the TOPPE MNS GE interpreter and modular spiral sequences.

Conclusion

We have designed a modular 2D gradient echo spiral sequence supporting several sequence formats and a modular metabolic‐specific 3D balanced steady‐state free precession sequence for 13 C acquisition, as well as developed a GE interpreter with X‐nucleus capabilities. Our work paves the way for future multi‐site studies with acquisitions for X‐nuclei across MRI vendors and software versions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedHumansPhantoms, ImagingBrainAnimalsKidneyCarbon IsotopesAlgorithmsSoftware

Stichwörter der Autor:innen

X‐nuclear MRImulti‐vendor MRIopen sourcepulse sequence programmingvendor‐neutral

Substanzen

Carbon Isotopes

Autor:innen und Institutionen

  • Department of Radiology and Biomedical Imaging, University of California, California, USA.
  • Department of Diagnostic Radiology and Nuclear Medicine, School of Medicine, University of Maryland, Baltimore, Maryland, USA.
  • Fischell Department of Bioengineering, University of Maryland, College Park, Maryland, USA.
  • Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • German Cancer Consortium (DKTK), Core Center Heidelberg and German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Functional MRI Laboratory, Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.

ORCID-Profile: Xiaoxi Liu, Di Cui, Peder E Z Larson, Jeremy W Gordon

Förderung

  • NCI NIH HHS
  • NIH HHS
  • UCSF Department of Radiology & Biomedical Imaging Seed Grant
  • Deutschen Konsortium für Translationale Krebsforschung
  • NINDS NIH HHS
  • NIBIB NIH HHS
  • NIA NIH HHS
  • NHLBI NIH HHS

Zitieren

APA

Liu, X., Cui, D., Larson, P. E. Z., Mayer, D., Korzowski, A., Nielsen, J. F., Schulte, R. F., Mu, C., Carvajal, L., Xu, D., Gordon, J. W., Vigneron, D. B., Flavell, R. R., & Wang, Z. J. (2025). Open‐source implementation of X‐nuclear sequences using the Pulseq framework. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30509

Vancouver

Liu X, Cui D, Larson PEZ, Mayer D, Korzowski A, Nielsen JF, et al. Open‐source implementation of X‐nuclear sequences using the Pulseq framework. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30509

BibTeX
@article{Liu2025,
  author = {Liu, Xiaoxi and Cui, Di and Larson, Peder E Z and Mayer, Dirk and Korzowski, Andreas and Nielsen, Jon-Fredrik and Schulte, Rolf F and Mu, Changhua and Carvajal, Lucas and Xu, Duan and Gordon, Jeremy W and Vigneron, Daniel B and Flavell, Robert R and Wang, Zhen J},
  title = {Open‐source implementation of X‐nuclear sequences using the Pulseq framework},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30509}
}
Kostenlos starten →