Submillimeter balanced SSFP BOLD–functional MRI accelerated with 3D stack‐of‐spirals at 9.4 T
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30064 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Praveen Iyyappan Valsala und 4 weiteren Autor:innen. Beteiligte Einrichtung: Magnetic Resonance Center, Max-Planck Institute for Biological Cybernetics, Tübingen, Germany. Hauptthemen laut MEDLINE: Algorithms, Magnetic Resonance Imaging, Imaging, Three-Dimensional, Brain. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Max-Planck-Society, Deutsche Forschungsgemeinschaft (the Reinhart Koselleck Project), European Research Council (Spread MRI).
Abstract (Originalsprache)
Purpose
This work aims to improve the speed of balanced SSFP (bSSFP) acquisition with segmented 3D stack‐of‐spirals for functional brain studies at ultrahigh field.
Methods
Functional experiments were performed with an accelerated 3D stack‐of‐spirals sequence with water excitation for fat suppression. The resulting data were reconstructed using an iterative algorithm with corrections for system imperfections such as trajectory deviations and B 0 inhomogeneity. In the first set of experiments, we evaluated the signal change and stability with respect to echo and TR for a full‐field checkerboard stimulus. To demonstrate the high spatio‐temporal resolution of the developed method, the results of three optimized protocols at submillimeter resolution (0.6‐mm isotropic and 0.8‐mm isotropic) and at 1.2 mm isotropic resolution for whole‐brain coverage were shown.
Results
Water excitation and the model‐based iterative reconstruction improved image quality. The BOLD‐related signal changes increased with longer TE and longer TR. We observed an increase in thermal noise performance at lower TE and higher TR. However, signal stability deteriorates at higher TE and TR. Therefore, optimized protocols used shorter TE and moderately long TR to maximize the sensitivity and speed. Reproducible activations were detected along the gray‐matter gyri in the submillimeter protocols with a median signal change of approximately 4% across subjects.
Conclusions
Three‐dimensional stack‐of‐spirals enables passband balanced SSFP functional imaging at a much higher spatial and temporal scale, compared with conventional spoiled gradient‐echo train sequences.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Magnetic Resonance Center, Max-Planck Institute for Biological Cybernetics, Tübingen, Germany.
- German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany.
- Department of Biomedical Magnetic Resonance, Eberhard Karls University Tübingen, Tübingen, Germany.
ORCID-Profile: Praveen Iyyappan Valsala, Marten Veldmann, Dario Bosch, Klaus Scheffler, Philipp Ehses
Förderung
- Max-Planck-Society
- Deutsche Forschungsgemeinschaft (the Reinhart Koselleck Project)
- European Research Council (Spread MRI)
- Bundesministerium für Bildung und Forschung
Zitieren
APA
Iyyappan Valsala, P., Veldmann, M., Bosch, D., Scheffler, K., & Ehses, P. (2024). Submillimeter balanced SSFP BOLD–functional MRI accelerated with 3D stack‐of‐spirals at 9.4 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30064
Vancouver
Iyyappan Valsala P, Veldmann M, Bosch D, Scheffler K, Ehses P. Submillimeter balanced SSFP BOLD–functional MRI accelerated with 3D stack‐of‐spirals at 9.4 T. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30064
BibTeX
@article{IyyappanValsala2024,
author = {Iyyappan Valsala, Praveen and Veldmann, Marten and Bosch, Dario and Scheffler, Klaus and Ehses, Philipp},
title = {Submillimeter balanced SSFP BOLD–functional MRI accelerated with 3D stack‐of‐spirals at 9.4 T},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30064}
}
EUVIMED