Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Jia Guo. Beteiligte Einrichtung: Department of Bioengineering, University of California Riverside, Riverside, California, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Angiography, White Matter. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NCI NIH HHS, NIBIB NIH HHS, NIH HHS.
Abstract (Originalsprache)
Purpose
Background suppression (BS) is recommended in arterial spin labeling (ASL) for improved SNR but is difficult to optimize in existing velocity‐selective ASL (VSASL) methods. Dual‐module VSASL (dm‐VSASL) enables delay‐insensitive, robust, and SNR‐efficient perfusion imaging, while allowing efficient BS, but its optimization has yet to be thoroughly investigated.
Methods
The inversion effects of the velocity‐selective labeling pulses, such as velocity‐selective inversion (VSI), can be used for BS, and were modeled for optimizing BS in dm‐VSASL. In vivo experiments using dual‐module VSI (dm‐VSI) were performed to compare two BS strategies: a conventional one with additional BS pulses and a new one without any BS pulse. Their BS performance, temporal noise, and temporal SNR were examined and compared, with pulsed and pseudo‐continuous ASL (PASL and PCASL) as the reference.
Results
The in vivo experiments validated the BS modeling. Strong positive linear correlations ( r > 0.82, p < 0.0001) between the temporal noise and the tissue signal were found in PASL/PCASL and dm‐VSI. Optimal BS can be achieved with and without additional BS pulses in dm‐VSI; the latter improved the ASL signals by 8.5% in gray matter ( p = 0.006) and 12.2% in white matter ( p = 0.014) and tended to provide better temporal SNR. The dm‐VSI measured significantly higher ASL signal ( p < 0.016) and temporal SNR ( p < 0.018) than PASL and PCASL. Complex reconstruction was found necessary with aggressive BS.
Conclusion
Guided by modeling, optimal BS can be achieved without any BS pulse in dm‐VSASL, further improving the ASL signal and the SNR performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Bioengineering, University of California Riverside, Riverside, California, USA.
ORCID-Profile: Jia Guo
Förderung
- NCI NIH HHS
- NIBIB NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Guo, J. (2024). Optimizing background suppression for dual‐module velocity‐selective arterial spin labeling: Without using additional background‐suppression pulses. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29995
Vancouver
Guo J. Optimizing background suppression for dual‐module velocity‐selective arterial spin labeling: Without using additional background‐suppression pulses. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29995
BibTeX
@article{Guo2024,
author = {Guo, Jia},
title = {Optimizing background suppression for dual‐module velocity‐selective arterial spin labeling: Without using additional background‐suppression pulses},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.29995}
}
EUVIMED