EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A mathematical model for velocity‐selective arterial spin labeling

Thomas T Liu iD, Eric C Wong, Divya S Bolar, Conan Chen, Ryan S Barnes

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.29935 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Thomas T Liu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Center for Functional MRI, University of California San Diego, La Jolla, California, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Angiography, Arteries. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To present a theoretical framework that rigorously defines and analyzes key concepts and quantities for velocity selective arterial spin labeling (VSASL). Theory and Methods An expression for the VSASL arterial delivery function is derived based on (1) labeling and saturation profiles as a function of velocity and (2) physiologically plausible approximations of changes in acceleration and velocity across the vascular system. The dependence of labeling efficiency on the amplitude and effective bolus width of the arterial delivery function is defined. Factors that affect the effective bolus width are examined, and timing requirements to minimize quantitation errors are derived.

Results

The model predicts that a flow‐dependent negative bias in the effective bolus width can occur when velocity selective inversion (VSI) is used for the labeling module and velocity selective saturation (VSS) is used for the vascular crushing module. The bias can be minimized by choosing a nominal labeling cutoff velocity that is lower than the nominal cutoff velocity of the vascular crushing module.

Conclusion

The elements of the model are specified in a general fashion such that future advances can be readily integrated. The model can facilitate further efforts to understand and characterize the performance of VSASL and provide critical theoretical insights that can be used to design future experiments and develop novel VSASL approaches.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance AngiographyArteriesSpin LabelsModels, TheoreticalAccelerationCerebrovascular CirculationBlood Flow Velocity

Stichwörter der Autor:innen

accelerationarterial spin labelingvelocity selective

Substanzen

Spin Labels

Autor:innen und Institutionen

  • Center for Functional MRI, University of California San Diego, La Jolla, California, USA.
  • Department of Radiology, University of California San Diego, La Jolla, California, USA.
  • Department of Psychiatry, University of California San Diego, La Jolla, California, USA.
  • Department of Electrical and Computer Engineering, University of California San Diego, La Jolla, California, USA.

ORCID-Profile: Thomas T Liu

Zitieren

APA

Liu, T. T., Wong, E. C., Bolar, D. S., Chen, C., & Barnes, R. S. (2024). A mathematical model for velocity‐selective arterial spin labeling. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29935

Vancouver

Liu TT, Wong EC, Bolar DS, Chen C, Barnes RS. A mathematical model for velocity‐selective arterial spin labeling. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29935

BibTeX
@article{Liu2024,
  author = {Liu, Thomas T and Wong, Eric C and Bolar, Divya S and Chen, Conan and Barnes, Ryan S},
  title = {A mathematical model for velocity‐selective arterial spin labeling},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.29935}
}
Kostenlos starten →