Dynamic B 0 field shimming for improving pseudo‐continuous arterial spin labeling at 7 T
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30387 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Yang Ji und 3 weiteren Autor:innen. Beteiligte Einrichtung: Wellcome Centre for Integrative Neuroimaging, FMRIB Division, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK. Hauptthemen laut MEDLINE: Brain, Image Processing, Computer-Assisted, Magnetic Resonance Imaging. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, National Natural Science Foundation of China.
Abstract (Originalsprache)
Purpose
B 0 field inhomogeneity within the brain‐feeding arteries is a major issue for pseudo‐continuous arterial spin labeling (PCASL) at 7 T because it reduces the labeling efficiency and leads to a loss of perfusion signal. This study aimed to develop a vessel‐specific dynamic B 0 field shimming method for 7 T PCASL to improve the labeling efficiency by correcting off‐resonance within the arteries in the labeling region.
Methods
We implemented a PCASL sequence with dynamic B 0 shimming at 7 T that compensates for B 0 field offsets in the brain‐feeding arteries by updating linear shimming terms and adding a phase increment to the PCASL RF pulses. Rapidly acquired vessel‐specific B 0 field maps were used to calculate dynamic B 0 shimming parameters. We evaluated both 2D and 3D variants of our method, comparing their performance against the established global frequency offset and optimal encoding scheme‐based corrections. Cerebral blood flow (CBF) maps were quantified before and after corrections, and CBF values from different methods were compared across the whole brain, white matter, and gray matter regions.
Results
All off‐resonance correction methods significantly recovered perfusion signals across the brain. The proposed vessel‐specific dynamic B 0 shimming method improved the labeling efficiency while maintaining optimal static shimming in the imaging region. Perfusion‐weighted images demonstrated the superiority of the 3D dynamic B 0 shimming method compared to global or 2D‐based correction approaches. CBF analysis revealed that 3D dynamic B 0 shimming significantly increased CBF values relative to the other methods.
Conclusion
Our proposed dynamic B 0 shimming method offers a significant advancement in PCASL robustness and effectiveness, enabling full utilization of 7 T ASL high sensitivity and spatial resolution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Wellcome Centre for Integrative Neuroimaging, FMRIB Division, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
- Department of Electronic Engineering and Information Science, School of Information Science and Technology, University of Science and Technology of China, Hefei, People's Republic of China.
- Institute of Science and Technology for Brain-inspired Intelligence, Fudan University, Shanghai, People's Republic of China.
ORCID-Profile: Yang Ji, Joseph G Woods, Hongwei Li, Thomas W Okell
Förderung
- Wellcome Trust
- National Natural Science Foundation of China
Zitieren
APA
Ji, Y., Woods, J. G., Li, H., & Okell, T. W. (2025). Dynamic B 0 field shimming for improving pseudo‐continuous arterial spin labeling at 7 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30387
Vancouver
Ji Y, Woods JG, Li H, Okell TW. Dynamic B 0 field shimming for improving pseudo‐continuous arterial spin labeling at 7 T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30387