Comparison of distortion correction preprocessing pipelines for DTI in the upper limb
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.29881 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Ryckie G Wade und 10 weiteren Autor:innen. Beteiligte Einrichtung: Leeds Institute for Medical Research, University of Leeds, Leeds, UK. Hauptthemen laut MEDLINE: Diffusion Tensor Imaging, Diffusion Magnetic Resonance Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Arthritis Research UK, British Heart Foundation, Medical Research Council.
Abstract (Originalsprache)
Purpose
DTI characterizes tissue microstructure and provides proxy measures of nerve health. Echo‐planar imaging is a popular method of acquiring DTI but is susceptible to various artifacts (e.g., susceptibility, motion, and eddy currents), which may be ameliorated via preprocessing. There are many pipelines available but limited data comparing their performance, which provides the rationale for this study.
Methods
DTI was acquired from the upper limb of heathy volunteers at 3T in blip‐up and blip‐down directions. Data were independently corrected using (i) FSL's TOPUP & eddy, (ii) FSL's TOPUP, (iii) DSI Studio, and (iv) TORTOISE. DTI metrics were extracted from the median, radial, and ulnar nerves and compared (between pipelines) using mixed‐effects linear regression. The geometric similarity of corrected b = 0 images and the slice matched T1‐weighted (T1w) images were computed using the Sörenson‐Dice coefficient.
Results
Without preprocessing, the similarity coefficient of the blip‐up and blip‐down datasets to the T1w was 0·80 and 0·79, respectively. Preprocessing improved the geometric similarity by 1% with no difference between pipelines. Compared to TOPUP & eddy, DSI Studio and TORTOISE generated 2% and 6% lower estimates of fractional anisotropy, and 6% and 13% higher estimates of radial diffusivity, respectively. Estimates of anisotropy from TOPUP & eddy versus TOPUP were not different but TOPUP reduced radial diffusivity by 3%. The agreement of DTI metrics between pipelines was poor.
Conclusions
Preprocessing DTI from the upper limb improves geometric similarity but the choice of the pipeline introduces clinically important variability in diffusion parameter estimates from peripheral nerves.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Leeds Institute for Medical Research, University of Leeds, Leeds, UK.
- Department of Plastic, Reconstructive and Hand Surgery, Leeds Teaching Hospitals Trust, Leeds, UK.
- Division of Psychology, Communication & Human Neuroscience, The University of Manchester, Manchester, UK.
- Leeds Teaching Hospitals Trust, Leeds, UK.
- Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
ORCID-Profile: Ryckie G Wade, Winnie Tam, Antonia Perumal, Sophanit Pepple, Timothy T Griffiths, Robert Flather, Hamied A Haroon, Sven Plein, Grainne Bourke, Irvin Teh
Förderung
- Arthritis Research UK
- British Heart Foundation
- Medical Research Council
- Department of Health and Social Care
- Medical Research Council Canada
- National Institute for Health and Care Research
- Leeds Biomedical Research Centre
Zitieren
APA
Wade, R. G., Tam, W., Perumal, A., Pepple, S., Griffiths, T. T., Flather, R., Haroon, H. A., Shelley, D., Plein, S., Bourke, G., & Teh, I. (2024). Comparison of distortion correction preprocessing pipelines for DTI in the upper limb. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29881
Vancouver
Wade RG, Tam W, Perumal A, Pepple S, Griffiths TT, Flather R, et al. Comparison of distortion correction preprocessing pipelines for DTI in the upper limb. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29881
BibTeX
@article{Wade2024,
author = {Wade, Ryckie G and Tam, Winnie and Perumal, Antonia and Pepple, Sophanit and Griffiths, Timothy T and Flather, Robert and Haroon, Hamied A and Shelley, David and Plein, Sven and Bourke, Grainne and Teh, Irvin},
title = {Comparison of distortion correction preprocessing pipelines for DTI in the upper limb},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.29881}
}
EUVIMED