Metal artifact reduction around cervical spine implant using diffusion tensor imaging at 3T: A phantom study

Slimane Tounekti, Mahdi Alizadeh, Devon Middleton, James S Harrop, Bassem Hiba, Laura Krisa, Choukri Mekkaoui, Feroze B Mohamed

Magnetic Resonance Imaging · 2024-01 · 105 · 57-66

DOI: 10.1016/j.mri.2023.11.007 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2024-01. Verfasst von Slimane Tounekti und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Thomas Jefferson University, Philadelphia, PA, USA. Electronic address: Slimane.tounekti@jefferson.edu. Hauptthemen laut MEDLINE: Diffusion Tensor Imaging, Artifacts. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NINDS NIH HHS, National Institute of Neurological Disorders and Stroke, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Diffusion MRI continues to play a key role in non-invasively assessing spinal cord integrity and pre-operative injury evaluation. However, post-operative Diffusion Tensor Imaging (DTI) acquisition of patients with metal implants results in severe geometric distortion. We propose and demonstrate a method to alleviate the technical challenges facing the acquisition of DTI on post-operative cases and longitudinal evaluation of therapeutics. MATERIAL AND METHODS: The described technique is based on the combination of the reduced Field-Of-View (rFOV) strategy and the phase segmented EPI, termed rFOV-PS-EPI. A custom-built phantom based on a cervical spine model with metal implants was used to collect DTI data at 3 Tesla scanner using: rFOV-PS-EPI, reduced Field-Of-View single-shot EPI (rFOV-SS-EPI), and conventional full FOV techniques including SS-EPI, PS-EPI, and readout-segmented EPI (RS-EPI). Geometric distortion, SNR, and signal void were assessed to evaluate images and compare the sequences. A two-sample t-test was performed with p-value of 0.05 or less to indicate statistical significance. RESULTS: The reduced FOV techniques showed better capability to reduce distortions compared to the Full FOV techniques. The rFOV-PS-EPI method provided DTI images of the phantom at the level of the hardware whereas the conventional rFOV-SS-EPI is useful only when the metal is approximately 20 mm away. In addition, compared to the rFOV-SS-EPI technique, the suggested approach produced smaller signal voids area as well as significantly reduced geometric distortion in Circularity (p 0.05). CONCLUSION: The combination of rFOV and a phase-segmented acquisition approach is effective for reducing metal-induced distortions in DTI scan on spinal cord with metal hardware at 3 T.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diffusion Tensor ImagingArtifactsHumansDiffusion Magnetic Resonance ImagingSpinal CordEcho-Planar ImagingCervical Vertebrae

Stichwörter der Autor:innen

3 TeslaDiffusion tensor imagingGeometric distortionMetal artifactsMetal implant

Autor:innen und Institutionen

  • Department of Radiology, Thomas Jefferson University, Philadelphia, PA, USA. Electronic address: Slimane.tounekti@jefferson.edu.
  • Department of Neurosurgery, Thomas Jefferson University, Philadelphia, PA, USA.
  • Department of Radiology, Thomas Jefferson University, Philadelphia, PA, USA.
  • Institut des Sciences Cognitives, CNRS UMR 5229, Université Lyon 1, Lyon, France.
  • Department of Physical Therapy, Thomas Jefferson University, Philadelphia, PA, USA.
  • Harvard Medical School, Boston, MA, USA; Department of Radiology, Massachusetts General Hospital, Boston, MA, USA; A.A. Martinos Center for Biomedical Imaging, Charlestown, MA, USA.

Förderung

  • NINDS NIH HHS
  • National Institute of Neurological Disorders and Stroke
  • National Institutes of Health

Zitieren

APA

Tounekti, S., Alizadeh, M., Middleton, D., Harrop, J. S., Hiba, B., Krisa, L., Mekkaoui, C., & Mohamed, F. B. (2024). Metal artifact reduction around cervical spine implant using diffusion tensor imaging at 3T: A phantom study. Magnetic Resonance Imaging, 105, 57-66. https://doi.org/10.1016/j.mri.2023.11.007

Vancouver

Tounekti S, Alizadeh M, Middleton D, Harrop JS, Hiba B, Krisa L, et al. Metal artifact reduction around cervical spine implant using diffusion tensor imaging at 3T: A phantom study. Magnetic Resonance Imaging. 2024;105:57-66. doi:10.1016/j.mri.2023.11.007

BibTeX
@article{Tounekti2024,
  author = {Tounekti, Slimane and Alizadeh, Mahdi and Middleton, Devon and Harrop, James S and Hiba, Bassem and Krisa, Laura and Mekkaoui, Choukri and Mohamed, Feroze B},
  title = {Metal artifact reduction around cervical spine implant using diffusion tensor imaging at 3T: A phantom study},
  journal = {Magnetic Resonance Imaging},
  year = {2024},
  volume = {105},
  pages = {57--66},
  doi = {10.1016/j.mri.2023.11.007}
}
Kostenlos starten →