Multi ‐ echo–based fat artifact correction for CEST MRI at 7 T
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.29863 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Katharina Tkotz und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Hauptthemen laut MEDLINE: Artifacts, Magnetic Resonance Imaging. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: FAU Emerging Fields Initiative, German Federal Ministry of Education and Research (BMBF).
Abstract (Originalsprache)
Purpose
CEST MRI is influenced by fat signal, which can reduce the apparent CEST contrast or lead to pseudo‐CEST effects. Our goal was to develop a fat artifact correction based on multi‐echo fat–water separation that functions stably for 7 T knee MRI data.
Methods
Our proposed algorithm utilizes the full complex data and a phase demodulation with an off‐resonance map estimation based on the Z‐spectra prior to fat–water separation to achieve stable fat artifact correction. Our method was validated and compared to multi‐echo–based methods originally proposed for 3 T by Bloch–McConnell simulations and phantom measurements. Moreover, the method was applied to in vivo 7 T knee MRI examinations and compared to Gaussian fat saturation and a published single‐echo Z‐spectrum–based fat artifact correction method.
Results
Phase demodulation prior to fat–water separation reduced the occurrence of fat–water swaps. Utilizing the complex signal data led to more stable correction results than working with magnitude data, as was proposed for 3 T. Our approach reduced pseudo‐nuclear Overhauser effects compared to the other correction methods. Thus, the mean asymmetry contrast at 3.5 ppm in cartilage over five volunteers increased from −9.2% (uncorrected) and −10.6% (Z‐spectrum–based) to −1.5%. Results showed higher spatial stability than with the fat saturation pulse.
Conclusion
Our work demonstrates the feasibility of multi‐echo–based fat–water separation with an adaptive fat model for fat artifact correction for CEST MRI at 7 T. Our approach provided better fat artifact correction throughout the entire spectrum and image than the fat saturation pulse or Z‐spectrum–based correction method for both phantom and knee imaging results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
- Department of Neuroradiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
- Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
- Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
ORCID-Profile: Katharina Tkotz, Andrzej Liebert, Lena V Gast, Moritz Zaiss, Armin M Nagel
Förderung
- FAU Emerging Fields Initiative
- German Federal Ministry of Education and Research (BMBF)
Zitieren
APA
Tkotz, K., Liebert, A., Gast, L. V., Zeiger, P., Uder, M., Zaiss, M., & Nagel, A. M. (2024). Multi ‐ echo–based fat artifact correction for CEST MRI at 7 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29863
Vancouver
Tkotz K, Liebert A, Gast LV, Zeiger P, Uder M, Zaiss M, et al. Multi ‐ echo–based fat artifact correction for CEST MRI at 7 T. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29863
BibTeX
@article{Tkotz2024,
author = {Tkotz, Katharina and Liebert, Andrzej and Gast, Lena V and Zeiger, Paula and Uder, Michael and Zaiss, Moritz and Nagel, Armin M},
title = {Multi ‐ echo–based fat artifact correction for CEST MRI at 7 T},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.29863}
}
EUVIMED