Lung parenchyma transverse relaxation rates at 0.55 T
Magnetic Resonance in Medicine · 2023-04 · 89 (4) · 1522-1530
DOI: 10.1002/mrm.29541 ↗At a glance
Journal article in Magnetic Resonance in Medicine, published online on 2022-11-20. Written by Bochao Li and 3 co-authors. Institution: Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, California, Los Angeles, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Respiration. Cited 19 times so far (Crossref); 36 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: National Science Foundation of Sri Lanka.
Abstract (original language)
Purpose
To determine R 2 and transverse relaxation rates in healthy lung parenchyma at 0.55 T. This is important in that it informs the design and optimization of new imaging methods for 0.55T lung MRI.
Methods
Experiments were performed in 3 healthy adult volunteers on a prototype whole‐body 0.55T MRI, using a custom free‐breathing electrocardiogram‐triggered, single‐slice echo‐shifted multi‐echo spin echo (ES‐MCSE) pulse sequence with respiratory navigation. Transverse relaxation rates R 2 and and off‐resonance ∆ f were jointly estimated using nonlinear least‐squares estimation. These measurements were compared against R 2 estimates from T 2 ‐prepared balanced SSFP (T 2 ‐Prep bSSFP) and estimates from multi‐echo gradient echo, which are used widely but prone to error due to different subvoxel weighting.
Results
The mean R 2 and values of lung parenchyma obtained from ES‐MCSE were 17.3 ± 0.7 Hz and 127.5 ± 16.4 Hz (T 2 = 61.6 ± 1.7 ms; = 9.5 ms ± 1.6 ms), respectively. The off‐resonance estimates ranged from −60 to 30 Hz. The R 2 from T 2 ‐Prep bSSFP was 15.7 ± 1.7 Hz (T 2 = 68.6 ± 8.6 ms) and from multi‐echo gradient echo was 131.2 ± 30.4 Hz ( = 8.0 ± 2.5 ms). Paired t‐test indicated that there is a significant difference between the proposed and reference methods ( p < 0.05). The mean R 2 estimate from T 2 ‐Prep bSSFP was slightly smaller than that from ES‐MCSE, whereas the mean and estimates from ES‐MCSE and multi‐echo gradient echo were similar to each other across all subjects.
Conclusions
Joint estimation of transverse relaxation rates and off‐resonance is feasible at 0.55 T with a free‐breathing electrocardiogram‐gated and navigator‐gated ES‐MCSE sequence. At 0.55 T, the mean R 2 of 17.3 Hz is similar to the reported mean R 2 of 16.7 Hz at 1.5 T, but the mean of 127.5 Hz is about 5–10 times smaller than that reported at 1.5 T.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, California, Los Angeles, USA.
- Siemens Medical Solutions USA, Los Angeles, California, USA.
- Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, California, Los Angeles, USA.
ORCID profiles: Bochao Li, Nam G Lee, Sophia X Cui, Krishna S Nayak
Funding
- National Science Foundation of Sri Lanka
Cite
APA
Li, B., Lee, N. G., Cui, S. X., & Nayak, K. S. (2023). Lung parenchyma transverse relaxation rates at 0.55 T. Magnetic Resonance in Medicine, 89(4), 1522-1530. https://doi.org/10.1002/mrm.29541
Vancouver
Li B, Lee NG, Cui SX, Nayak KS. Lung parenchyma transverse relaxation rates at 0.55 T. Magnetic Resonance in Medicine. 2023;89(4):1522-1530. doi:10.1002/mrm.29541
BibTeX
@article{Li2023,
author = {Li, Bochao and Lee, Nam G and Cui, Sophia X and Nayak, Krishna S},
title = {Lung parenchyma transverse relaxation rates at 0.55 T},
journal = {Magnetic Resonance in Medicine},
year = {2023},
volume = {89},
number = {4},
pages = {1522--1530},
doi = {10.1002/mrm.29541}
}
EUVIMED