Lung parenchyma transverse relaxation rates at 0.55 T
Magnetic Resonance in Medicine · 2023-04 · 89 (4) · 1522-1530
DOI: 10.1002/mrm.29541 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 20.11.2022. Verfasst von Bochao Li und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, California, Los Angeles, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Respiration. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Science Foundation of Sri Lanka.
Abstract (Originalsprache)
Purpose
To determine R 2 and transverse relaxation rates in healthy lung parenchyma at 0.55 T. This is important in that it informs the design and optimization of new imaging methods for 0.55T lung MRI.
Methods
Experiments were performed in 3 healthy adult volunteers on a prototype whole‐body 0.55T MRI, using a custom free‐breathing electrocardiogram‐triggered, single‐slice echo‐shifted multi‐echo spin echo (ES‐MCSE) pulse sequence with respiratory navigation. Transverse relaxation rates R 2 and and off‐resonance ∆ f were jointly estimated using nonlinear least‐squares estimation. These measurements were compared against R 2 estimates from T 2 ‐prepared balanced SSFP (T 2 ‐Prep bSSFP) and estimates from multi‐echo gradient echo, which are used widely but prone to error due to different subvoxel weighting.
Results
The mean R 2 and values of lung parenchyma obtained from ES‐MCSE were 17.3 ± 0.7 Hz and 127.5 ± 16.4 Hz (T 2 = 61.6 ± 1.7 ms; = 9.5 ms ± 1.6 ms), respectively. The off‐resonance estimates ranged from −60 to 30 Hz. The R 2 from T 2 ‐Prep bSSFP was 15.7 ± 1.7 Hz (T 2 = 68.6 ± 8.6 ms) and from multi‐echo gradient echo was 131.2 ± 30.4 Hz ( = 8.0 ± 2.5 ms). Paired t‐test indicated that there is a significant difference between the proposed and reference methods ( p < 0.05). The mean R 2 estimate from T 2 ‐Prep bSSFP was slightly smaller than that from ES‐MCSE, whereas the mean and estimates from ES‐MCSE and multi‐echo gradient echo were similar to each other across all subjects.
Conclusions
Joint estimation of transverse relaxation rates and off‐resonance is feasible at 0.55 T with a free‐breathing electrocardiogram‐gated and navigator‐gated ES‐MCSE sequence. At 0.55 T, the mean R 2 of 17.3 Hz is similar to the reported mean R 2 of 16.7 Hz at 1.5 T, but the mean of 127.5 Hz is about 5–10 times smaller than that reported at 1.5 T.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, California, Los Angeles, USA.
- Siemens Medical Solutions USA, Los Angeles, California, USA.
- Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, California, Los Angeles, USA.
ORCID-Profile: Bochao Li, Nam G Lee, Sophia X Cui, Krishna S Nayak
Förderung
- National Science Foundation of Sri Lanka
Zitieren
APA
Li, B., Lee, N. G., Cui, S. X., & Nayak, K. S. (2023). Lung parenchyma transverse relaxation rates at 0.55 T. Magnetic Resonance in Medicine, 89(4), 1522-1530. https://doi.org/10.1002/mrm.29541
Vancouver
Li B, Lee NG, Cui SX, Nayak KS. Lung parenchyma transverse relaxation rates at 0.55 T. Magnetic Resonance in Medicine. 2023;89(4):1522-1530. doi:10.1002/mrm.29541
BibTeX
@article{Li2023,
author = {Li, Bochao and Lee, Nam G and Cui, Sophia X and Nayak, Krishna S},
title = {Lung parenchyma transverse relaxation rates at 0.55 T},
journal = {Magnetic Resonance in Medicine},
year = {2023},
volume = {89},
number = {4},
pages = {1522--1530},
doi = {10.1002/mrm.29541}
}