Tensor denoising of multidimensional MRI data
Magnetic Resonance in Medicine · 2023-03 · 89 (3) · 1160-1172
DOI: 10.1002/mrm.29478 ↗At a glance
Journal article in Magnetic Resonance in Medicine, published online on 2022-10-11. Written by Jonas L Olesen and 4 co-authors. Institution: Center of Functionally Integrative Neuroscience, Department of Clinical Medicine, Aarhus University, Aarhus, Denmark. Major MEDLINE topics: Algorithms, Magnetic Resonance Imaging. Cited 84 times so far (Crossref); 41 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: NIBIB NIH HHS, “la Caixa” Foundation, Danmarks Grundforskningsfond.
Abstract (original language)
Purpose
To develop a denoising strategy leveraging redundancy in high‐dimensional data. Theory and Methods The SNR fundamentally limits the information accessible by MRI. This limitation has been addressed by a host of denoising techniques, recently including the so‐called MPPCA: principal component analysis of the signal followed by automated rank estimation, exploiting the Marchenko‐Pastur distribution of noise singular values. Operating on matrices comprised of data patches, this popular approach objectively identifies noise components and, ideally, allows noise to be removed without introducing artifacts such as image blurring, or nonlocal averaging. The MPPCA rank estimation, however, relies on a large number of noise singular values relative to the number of signal components to avoid such ill effects. This condition is unlikely to be met when data patches and therefore matrices are small, for example due to spatially varying noise. Here, we introduce tensor MPPCA (tMPPCA) for the purpose of denoising multidimensional data, such as from multicontrast acquisitions. Rather than combining dimensions in matrices, tMPPCA uses each dimension of the multidimensional data's inherent tensor‐structure to better characterize noise, and to recursively estimate signal components.
Results
Relative to matrix‐based MPPCA, tMPPCA requires no additional assumptions, and comparing the two in a numerical phantom and a multi‐TE diffusion MRI data set, tMPPCA dramatically improves denoising performance. This is particularly true for small data patches, suggesting that tMPPCA can be especially beneficial in such cases.
Conclusions
The MPPCA denoising technique can be extended to high‐dimensional data with improved performance for smaller patch sizes.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Center of Functionally Integrative Neuroscience, Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.
- Department of Physics and Astronomy, Aarhus University, Aarhus, Denmark.
- Champalimaud Research, Champalimaud Foundation, Lisbon, Portugal.
ORCID profiles: Jonas L Olesen, Andrada Ianus, Leif Østergaard, Noam Shemesh, Sune N Jespersen
Funding
- NIBIB NIH HHS
- “la Caixa” Foundation
- Danmarks Grundforskningsfond
- European Regional Development Fund
- H2020 European Research Council
- Velux Fonden
Cite
APA
Olesen, J. L., Ianus, A., Østergaard, L., Shemesh, N., & Jespersen, S. N. (2023). Tensor denoising of multidimensional MRI data. Magnetic Resonance in Medicine, 89(3), 1160-1172. https://doi.org/10.1002/mrm.29478
Vancouver
Olesen JL, Ianus A, Østergaard L, Shemesh N, Jespersen SN. Tensor denoising of multidimensional MRI data. Magnetic Resonance in Medicine. 2023;89(3):1160-1172. doi:10.1002/mrm.29478
BibTeX
@article{Olesen2023,
author = {Olesen, Jonas L and Ianus, Andrada and Østergaard, Leif and Shemesh, Noam and Jespersen, Sune N},
title = {Tensor denoising of multidimensional MRI data},
journal = {Magnetic Resonance in Medicine},
year = {2023},
volume = {89},
number = {3},
pages = {1160--1172},
doi = {10.1002/mrm.29478}
}