Independent verification of vendor‐issued dosimetric data for 106 Ru brachytherapy using diode detectors with traceability to external beam standards for absorbed dose to water

Simon Dahlander, Ilias Billas, Thorsten Sander, Graham Bass, Linda Persson, Åsa Carlsson Tedgren

Medical Physics · 2026-04 · 53 (4)

DOI: 10.1002/mp.70381 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 29.03.2026. Verfasst von Simon Dahlander und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology Pathology, Karolinska Institute, Stockholm, Sweden. Hauptthemen laut MEDLINE: Brachytherapy, Water, Ruthenium Radioisotopes, Radiometry. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swedish Radiation Safety Authority, Swedish Cancer Foundation, Strålsäkerhetsmyndigheten.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Independent verification of the manufacturer‐provided dosimetry data for 106 Ru ophthalmic brachytherapy applicators is crucial for safe and accurate treatment, yet a standardized, traceable method for clinical absolute dosimetry has been lacking.

Purpose

This work establishes a complete framework for traceable absorbed dose to water measurements of 106 Ru eye plaques in absolute units of Gray, complementing the high‐precision BetaCheck‐106™ setup with a robust detector calibration methodology independent of the source manufacturer.

Methods

Three microSilicon diode detectors were calibrated in traceable 60 Co and 6 MeV electron reference beams. Depth‐dose measurements for four CCB‐type 106 Ru plaques were performed in water using the BetaCheck‐106™ setup. Monte Carlo (MC) simulations were employed to calculate depth‐dependent beam quality correction factors which account for the detector's response in the 106 Ru field relative to the calibration beams. The method was validated against measurements performed at the National Physical Laboratory (NPL) using alanine dosimetry. MC simulations were also used to investigate the water‐equivalence of the NPL alanine/PMMA phantom setup.

Results

The MC‐calculated correction factors for the diodes showed a significant depth‐dependence, underscoring the necessity of such corrections in the steep 106 Ru depth‐dose gradient. The dose rates determined with the calibrated diodes were in agreement with the NPL alanine results. Both methods yielded dose rates systematically lower than those provided in the manufacturer's certificates, though generally within the stated uncertainties. The MC simulations revealed substantial non‐water equivalence correction factors for the alanine/PMMA phantom, highlighting the advantage of direct measurements in water.

Conclusions

We present a novel, comprehensive methodology for independent and traceable absolute dosimetry of 106 Ru applicators. By combining a dedicated water phantom setup with diode detectors calibrated against external beam standards and MC‐derived correction factors, this framework empowers clinical users to perform robust verification measurements, filling a critical gap in the quality assurance of ocular brachytherapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyWaterRuthenium RadioisotopesRadiometryRadiation DosageMonte Carlo MethodCalibrationRadiotherapy DosageReference Standards

Stichwörter der Autor:innen

Monte CarloRu‐106 brachytherapydosimetrytraceable metrology

Substanzen

WaterRuthenium RadioisotopesRuthenium-106

Autor:innen und Institutionen

  • Department of Oncology Pathology, Karolinska Institute, Stockholm, Sweden.
  • Department of Nuclear Medicine and Medical Physics, Karolinska University Hospital, Stockholm, Sweden.
  • National Physical Laboratory, Teddington, UK.
  • Swedish Radiation Safety Authority, Stockholm, Sweden.
  • Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden.

Förderung

  • Swedish Radiation Safety Authority
  • Swedish Cancer Foundation
  • Strålsäkerhetsmyndigheten

Zitieren

APA

Dahlander, S., Billas, I., Sander, T., Bass, G., Persson, L., & Tedgren, Å. C. (2026). Independent verification of vendor‐issued dosimetric data for 106 Ru brachytherapy using diode detectors with traceability to external beam standards for absorbed dose to water. Medical Physics, 53(4). https://doi.org/10.1002/mp.70381

Vancouver

Dahlander S, Billas I, Sander T, Bass G, Persson L, Tedgren ÅC. Independent verification of vendor‐issued dosimetric data for 106 Ru brachytherapy using diode detectors with traceability to external beam standards for absorbed dose to water. Medical Physics. 2026;53(4). doi:10.1002/mp.70381

BibTeX
@article{Dahlander2026,
  author = {Dahlander, Simon and Billas, Ilias and Sander, Thorsten and Bass, Graham and Persson, Linda and Tedgren, Åsa Carlsson},
  title = {Independent verification of vendor‐issued dosimetric data for 106 Ru brachytherapy using diode detectors with traceability to external beam standards for absorbed dose to water},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {4},
  doi = {10.1002/mp.70381}
}
Kostenlos starten →