A dynamic anthropomorphic phantom for end‐to‐end testing in image‐ and surface‐guided adaptive radiotherapy
Medical Physics · 2025
DOI: 10.1002/mp.70107 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Anahita Bakhtiari Moghaddam und 18 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine, University of Heidelberg, Heidelberg, Germany. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiotherapy, Image-Guided. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Respiratory and digestive motion can compromise the accuracy of radiotherapy (RT), affecting tumor targeting and healthy tissue sparing. Current phantoms often fail to replicate complex organ movements and lack compatibility with advanced imaging modalities like magnetic resonance imaging (MRI) and 4D computed tomography (4DCT), limiting their utility in adaptive radiotherapy. The BRaVIDA (Breathing Radiotherapy Visual monitoring, Imaging, and Dosimetric Anthropomorphic) phantom addresses these limitations by simulating realistic organ motion during breathing and digestion. Its MRI compatibility allows for comprehensive testing in adaptive radiotherapy workflows, improving motion management and treatment precision for both photon and ion beam therapies.
Purpose
This study introduces the BRaVIDA ( B reathing Ra diotherapy V isual monitoring, I maging, and D osimetric A nthropomorphic) phantom, a novel anthropomorphic phantom designed to simulate respiratory motion and support end‐to‐end testing in adaptive radiotherapy. The aim is to improve motion management, imaging accuracy, and dosimetric validation for radiotherapy treatments.
Methods
The BRaVIDA phantom is designed with realistic anatomical structures of the thorax and abdomen and may be equipped with dosimeters. It simulates respiratory and digestive motion using an in‐house developed hydraulic system. The phantom was imaged by CT and MRI, using different protocols (CT, MRI, 4DCT, and 4DMRI) to evaluate image contrast and motion parameters. Dosimetric properties were assessed by measuring the relative electron density (RED) and the stopping power ratio (SPR) of ion beams and the results were compared with the values of the treatment planning system (TPS).
Results
The phantom demonstrated realistic image contrast in CT and MRI, with anthropomorphic characteristics similar to human tissues. Motion amplitudes for various organ models (pancreas, stomach, liver) were successfully measured using 4DCT and 4DMRI. Dosimetric testing showed that the RED and SPR values of BRaVIDA align closely with TPS reference values. Deviations in photon attenuation were below 1.5% for all phantom materials.
Conclusions
BRaVIDA presents a versatile, MRI‐compatible phantom exhibiting anthropomorphic image contrast, inter‐ and intrafractional motion as well as radiation attenuation in photon beams, and equipping the phantom with detectors allows for full end‐to‐end‐tests in adaptive image‐ and surface‐guided photon RT workflows without further adaptions. For ion beams, the SPR values in the TPS have to be adapted to the measured data to assure correct range calculation in the phantom. The phantom presents a valuable and accurate tool for clinical and research applications in adaptive RT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Medicine, University of Heidelberg, Heidelberg, Germany.
- Division of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- Heidelberg Institute for Radiation Oncology (HIRO), National Center for Radiation Research in Oncology (NCRO), Heidelberg, Germany.
- Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- Heidelberg Ion Beam Therapy Center (HIT), University Hospital Heidelberg, Heidelberg, Germany.
- Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
Zitieren
APA
Bakhtiari Moghaddam, A., Runz, A., Augusto, R. F., Echner, G., Johnen, W., Gabriel, R., Haering, P., Lang, C., Seeber, S., Murillo, C., Ackermann, B., Pestana, R., Beyer, C., Weykamp, F., Jochim, M., Qubala, A., Batista, V., Jäkel, O., & Karger, C. P. (2025). A dynamic anthropomorphic phantom for end‐to‐end testing in image‐ and surface‐guided adaptive radiotherapy. Medical Physics. https://doi.org/10.1002/mp.70107
Vancouver
Bakhtiari Moghaddam A, Runz A, Augusto RF, Echner G, Johnen W, Gabriel R, et al. A dynamic anthropomorphic phantom for end‐to‐end testing in image‐ and surface‐guided adaptive radiotherapy. Medical Physics. 2025. doi:10.1002/mp.70107
BibTeX
@article{BakhtiariMoghaddam2025,
author = {Bakhtiari Moghaddam, Anahita and Runz, A and Augusto, R Figueiredo and Echner, G and Johnen, W and Gabriel, R and Haering, P and Lang, C and Seeber, S and Murillo, C and Ackermann, B and Pestana, R and Beyer, C and Weykamp, F and Jochim, M and Qubala, A and Batista, V and Jäkel, O and Karger, C P},