Affordable realtime respiratory monitoring system for radiation oncology
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70250 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Mandeep Kaur und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiation Oncology, Radiotherapy Planning, Computer-Assisted, Respiration. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Winship Cancer Institute of Emory University, Winship Cancer Institute.
Abstract (Originalsprache)
Background and purpose Managing respiratory motion is essential for accurate external beam radiotherapy of thoracic and abdominal tumors. Surface‐guided radiation therapy systems offer precise respiratory motion monitoring, albeit being often expensive and complex. Cost‐effective optical imaging technology could increase accessibility to accurate respiratory monitoring in radiotherapy. Materials and methods A realtime respiratory monitoring system was developed using off‐the‐shelf depth camera and custom software. The system tracks a superficial surrogate in three dimensions using depth and optical data. Initial testing involved tracking rigid phantoms in static and dynamic scenarios. 1D sinusoidal motion of various amplitudes was used as ground truth to assess system performance at different distances from the camera. 3D‐motion was tracked concomitantly with the proposed system and a clinical surface imaging system. System accuracy against nominal motion was evaluated using Pearson correlation (r), tracking success rate (TSR), and mean absolute error (MAE). Bland–Altman analysis assessed agreement with the clinical system in the absence of ground truth.
Results
The system accurately measured static targets, with depth errors of 0.03 ± 0.13 mm at 560 mm and 0.65 ± 0.61 mm at 1120 mm. For vertical 1D‐motion, TSR (≤1 mm) was 100% at 560 mm and 98.7% at 1120 mm; longitudinal TSR ranged from 92% to 50% as distance increased. Pearson correlation remained > 0.98, except at distances ≥980 mm and amplitudes ≤5 mm. Mean errors were consistently ≤1 mm, with minimal bias and narrow limits of agreement in Bland–Altman analysis compared to the clinical system.
Conclusion
The proposed system demonstrates suitability for realtime respiratory motion monitoring in radiotherapy, while minimizing the implementation costs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia, USA.
ORCID-Profile: Mandeep Kaur, Marian Axente
Förderung
- Winship Cancer Institute of Emory University
- Winship Cancer Institute
Zitieren
APA
Kaur, M., Al-Hallaq, H., & Axente, M. (2025). Affordable realtime respiratory monitoring system for radiation oncology. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70250
Vancouver
Kaur M, Al-Hallaq H, Axente M. Affordable realtime respiratory monitoring system for radiation oncology. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70250
BibTeX
@article{Kaur2025,
author = {Kaur, Mandeep and Al-Hallaq, Hania and Axente, Marian},
title = {Affordable realtime respiratory monitoring system for radiation oncology},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70250}
}