EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Development and first implementation of a novel multi‐modality cardiac motion and dosimetry phantom for radiotherapy applications

Kenneth W Gregg, Chase Ruff, Grant Koenig, Kalin I Penev, Andrew Shepard, Grace Kreissler, Margo Amatuzio, Cameron Owens, Prashant Nagpal, Carri K Glide-Hurst

Medical Physics · 2024

DOI: 10.1002/mp.17315 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Kenneth W Gregg und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Human Oncology, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Heart, Radiometry, Movement. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NHLBI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Cardiac applications in radiation therapy are rapidly expanding including magnetic resonance guided radiation therapy (MRgRT) for real‐time gating for targeting and avoidance near the heart or treating ventricular tachycardia (VT).

Purpose

This work describes the development and implementation of a novel multi‐modality and magnetic resonance (MR)‐compatible cardiac phantom.

Methods

The patient‐informed 3D model was derived from manual contouring of a contrast‐enhanced Coronary Computed Tomography Angiography scan, exported as a Stereolithography model, then post‐processed to simulate female heart with an average volume. The model was 3D‐printed using Elastic50A to provide MR contrast to water background. Two rigid acrylic modules containing cardiac structures were designed and assembled, retrofitting to an MR‐safe programmable motor to supply cardiac and respiratory motion in superior‐inferior directions. One module contained a cavity for an ion chamber (IC), and the other was equipped with multiple interchangeable cavities for plastic scintillation detectors (PSDs). Images were acquired on a 0.35 T MR‐linac for validation of phantom geometry, motion, and simulated online treatment planning and delivery. Three motion profiles were prescribed: patient‐derived cardiac (sine waveform, 4.3 mm peak‐to‐peak, 60 beats/min), respiratory (cos 4 waveform, 30 mm peak‐to‐peak, 12 breaths/min), and a superposition of cardiac (sine waveform, 4 mm peak‐to‐peak, 70 beats/min) and respiratory (cos 4 waveform, 24 mm peak‐to‐peak, 12 breaths/min). The amplitude of the motion profiles was evaluated from sagittal cine images at eight frames/s with a resolution of 2.4 mm × 2.4 mm. Gated dosimetry experiments were performed using the two module configurations for calculating dose relative to stationary. A CT‐based VT treatment plan was delivered twice under cone‐beam CT guidance and cumulative stationary doses to multi‐point PSDs were evaluated.

Results

No artifacts were observed on any images acquired during phantom operation. Phantom excursions measured 49.3 ± 25.8%/66.9 ± 14.0%, 97.0 ± 2.2%/96.4 ± 1.7%, and 90.4 ± 4.8%/89.3 ± 3.5% of prescription for cardiac, respiratory, and cardio‐respiratory motion profiles for the 2‐chamber (PSD) and 12‐substructure (IC) phantom modules respectively. In the gated experiments, the cumulative dose was

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingHeartRadiometryMovementMagnetic Resonance ImagingRadiotherapy, Image-GuidedHumansFemale

Stichwörter der Autor:innen

MRgRTcardiac phantomdosimetrymulti‐modality

Autor:innen und Institutionen

  • Department of Human Oncology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Modus Medical Devices, Inc. (IBA QUASAR), London, Ontario, Canada.
  • Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.

Förderung

  • NIBIB NIH HHS
  • NHLBI NIH HHS

Zitieren

APA

Gregg, K. W., Ruff, C., Koenig, G., Penev, K. I., Shepard, A., Kreissler, G., Amatuzio, M., Owens, C., Nagpal, P., & Glide-Hurst, C. K. (2024). Development and first implementation of a novel multi‐modality cardiac motion and dosimetry phantom for radiotherapy applications. Medical Physics. https://doi.org/10.1002/mp.17315

Vancouver

Gregg KW, Ruff C, Koenig G, Penev KI, Shepard A, Kreissler G, et al. Development and first implementation of a novel multi‐modality cardiac motion and dosimetry phantom for radiotherapy applications. Medical Physics. 2024. doi:10.1002/mp.17315

BibTeX
@article{Gregg2024,
  author = {Gregg, Kenneth W and Ruff, Chase and Koenig, Grant and Penev, Kalin I and Shepard, Andrew and Kreissler, Grace and Amatuzio, Margo and Owens, Cameron and Nagpal, Prashant and Glide-Hurst, Carri K},
  title = {Development and first implementation of a novel multi‐modality cardiac motion and dosimetry phantom for radiotherapy applications},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17315}
}