Redundancy‐weighted FDK reconstruction for dual‐detector combined‐scanning CBCT: Practical implementation for image guided particle therapy
Medical Physics · 2025-08 · 52 (8)
DOI: 10.1002/mp.17996 ↗At a glance
Journal article in Medical Physics, published online on 2025-08-08. Written by Bongyong Song and 14 co-authors. Institution: Department of Radiation Oncology, University of California San Diego, San Diego, California, USA. Major MEDLINE topics: Cone-Beam Computed Tomography, Radiotherapy, Image-Guided, Proton Therapy. Cited 2 times so far (Crossref); 49 references. Funded by: National Cancer Center (NCC), State of Florida Casey DeSantis funds.
Abstract (original language)
Background
Cone‐beam computed tomography (CBCT) is essential for image‐guided particle therapy (IGPT), providing daily patient positioning, anatomical monitoring, and treatment verification. However, conventional single‐detector CBCT suffers from poor soft‐tissue contrast, long acquisition times, and motion artifacts, reducing its effectiveness for adaptive radiotherapy workflows. Addressing these challenges requires an advanced CBCT reconstruction approach capable of enhancing image quality while reducing scan time and radiation exposure.
Purpose
This study introduces a Dual Detector Combined Scanning (DDCS) CBCT reconstruction algorithm to overcome conventional CBCT limitations. By integrating a dual‐source, orthogonal imaging setup, DDCS significantly reduces scanning time and imaging artifacts. A modified Parker‐weighting algorithm further improves image reconstruction accuracy, ensuring high‐fidelity visualization. The goal is to enhance adaptive radiotherapy by improving real‐time patient positioning, dose verification, and motion management.
Methods
A dual‐detector CBCT system was designed to acquire orthogonal projections simultaneously, reducing scan time and improving data completeness. A modified Parker‐weighting algorithm corrected angular overlap distortions, ensuring better reconstruction accuracy. The system was evaluated using numerical and physical phantom studies to assess spatial resolution, contrast‐to‐noise ratio (CNR), and artifact reduction. Additionally, a dynamic phantom simulated respiratory motion, validating motion robustness for adaptive radiotherapy. Performance was compared against single‐detector CBCT, focusing on image fidelity, noise suppression, and computational efficiency.
Results
DDCS‐CBCT demonstrated higher CNR, reduced motion artifacts, and improved spatial resolution, leading to more accurate anatomical visualization. The system's efficiency enables faster, more reliable patient setup in IGPT while maintaining a lower imaging dose.
Conclusion
The proposed DDCS‐CBCT approach significantly improves imaging accuracy, reduces scan time, and enhances real‐time volumetric guidance in IGRT/IGPT. These findings support the clinical feasibility of the DDCS system and its integration into online adaptive radiotherapy workflows.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Radiation Oncology, University of California San Diego, San Diego, California, USA.
- Department of Radiation Oncology, MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, Newyork, New York, USA.
- Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.
- Proton Therapy Center, National Cancer Center Korea, Goyang-si, South Korea.
- Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University College of Medicine, Seoul, South Korea.
Funding
- National Cancer Center (NCC)
- State of Florida Casey DeSantis funds
Cite
APA
Song, B., Whitaker, T. J., Gautam, A. S., Olberg, S., Choi, B., Lee, S. U., Jeong, J. H., Kim, J., Liang, X., Tan, J., Lu, B., Yaddanapudi, S., Furutani, K., Beltran, C. J., & Park, J. C. (2025). Redundancy‐weighted FDK reconstruction for dual‐detector combined‐scanning CBCT: Practical implementation for image guided particle therapy. Medical Physics, 52(8). https://doi.org/10.1002/mp.17996
Vancouver
Song B, Whitaker TJ, Gautam AS, Olberg S, Choi B, Lee SU, et al. Redundancy‐weighted FDK reconstruction for dual‐detector combined‐scanning CBCT: Practical implementation for image guided particle therapy. Medical Physics. 2025;52(8). doi:10.1002/mp.17996
BibTeX
@article{Song2025,
author = {Song, Bongyong and Whitaker, Thomas J and Gautam, Archana S and Olberg, Sven and Choi, Byongsu and Lee, Sung Uk and Jeong, Jong Hwi and Kim, Jeongheon and Liang, Xiaoying and Tan, Jun and Lu, Bo and Yaddanapudi, Sridhar and Furutani, Keith and Beltran, Chris J and Park, Justin C},
title = {Redundancy‐weighted FDK reconstruction for dual‐detector combined‐scanning CBCT: Practical implementation for image guided particle therapy},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {8},
doi = {10.1002/mp.17996}
}
EUVIMED