The implementation of an image‐guided system at a proton therapy center facility
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14181 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Bijan Arjomandy und 7 weiteren Autor:innen. Beteiligte Einrichtung: Karmanos Cancer Institute at McLaren-Flint, McLaren Proton Therapy Center, Flint, Michigan, USA. Hauptthemen laut MEDLINE: Proton Therapy, Radiosurgery, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Pencil Beam Scanning (PBS) proton therapy has similar requirements on patient alignment to within 1 mm and 1‐degree accuracy as photon radiosurgery. This study describes general workflow, acceptance, and commissioning test procedures and their respective results for an independent robotic arm used for Image Guided Radiotherapy (IGRT) for a Proton Therapy System.
Methods
The system is equipped with kV‐imaging techniques capable of orthogonal and Cone‐Beam Computed Tomography (CBCT) imaging modalities mounted on an independent robotic arm gantry attached to the ceiling. The imaging system is capable of 360‐degree rotation around patients to produce CBCT and kilovoltage orthogonal images. The imaging hardware is controlled by Ehmet Health XIS software, and MIM Software handles the image fusion and registration to an acceptable accuracy of ≤1‐mm shifts for patients’ alignment. The system was tested according to the requirements outlined in the American Association of Physicists in Medicine (AAPM) Task Group (TG) 142 and TG 179. The system tests included (1) safety, functionality, and connectivity, (2) mechanical testing, (3) image quality, (4) image registration, and (5) imaging dose. Additional tests included imaging gantry isocentricity with a laser tracker and collision‐avoiding system checks.
Results
The orthogonal and volumetric imaging are comparable in quality to other commercially available On‐Board Imagers (OBI) systems. The resulting spatial resolution values were 1.8‐, 0.8‐, and 0.5‐Line Pairs per Millimeter (lp/mm) for orthogonal, full‐fan CBCT, and half‐fan CBCT, respectively. The image registration is accurate to within 1 mm and 1 degree. The data shows consistent imaging‐guided system performance with standard deviations in x, y, and z of 0.7, 0.8, and 0.7 mm, respectively.
Conclusions
The system provides excellent image quality and performance, which can be used for IGRT. The proven accuracy of the x‐ray imaging and positioning system at McLaren Proton Therapy Center (MPTC) is 1 mm, making it suitable for proton therapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Karmanos Cancer Institute at McLaren-Flint, McLaren Proton Therapy Center, Flint, Michigan, USA.
- Erlanger Baroness Hospital, Chattanooga, Tennessee, USA.
- Ehmet Health, Plymouth, Michigan, USA.
Zitieren
APA
Arjomandy, B., Bejarano Buele, A. I., Clinthorne, N., Vujasevic, M., Athar, B., Deemer, J., Alkhatib, A., & Hussain, A. (2024). The implementation of an image‐guided system at a proton therapy center facility. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14181
Vancouver
Arjomandy B, Bejarano Buele AI, Clinthorne N, Vujasevic M, Athar B, Deemer J, et al. The implementation of an image‐guided system at a proton therapy center facility. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14181
BibTeX
@article{Arjomandy2024,
author = {Arjomandy, Bijan and Bejarano Buele, Ana Isabel and Clinthorne, Neal and Vujasevic, Milos and Athar, Basit and Deemer, James and Alkhatib, Ahmad and Hussain, Abrar},
title = {The implementation of an image‐guided system at a proton therapy center facility},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14181}
}
EUVIMED