A Gantry-Mounted Photon-Counting Detector Computed Tomography Prototype for Image Guided Proton Therapy
International Journal of Radiation Oncology*Biology*Physics · 2026
DOI: 10.1016/j.ijrobp.2026.02.210 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2026. Verfasst von Christian De Caro und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin. Hauptthemen laut MEDLINE: Proton Therapy, Photons, Radiotherapy, Image-Guided, Cone-Beam Computed Tomography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 64 Literaturangaben. Gefördert von: NIBIB NIH HHS.
Abstract (Originalsprache)
PURPOSE: Currently, on-board kilovoltage imaging systems on proton therapy machines typically employ scintillator-based energy-integrating detectors (EIDs) to acquire planar images or cone beam computed tomography (CBCT) scans. However, EID-based CBCT is limited by suboptimal image quality and poor quantification accuracy, falling short of the requirements for online adaptive proton therapy and other advanced imaging tasks for proton therapy. This study aims to experimentally demonstrate the feasibility of on-board photon-counting detector (PCD) CT imaging by integrating a PCD with a proton therapy gantry. METHODS AND MATERIALS: A cadmium telluride (CdTe)-based PCD, featuring a pixel size of 100μm and 2 energy channels, was mounted onto the surface of the existing flat-panel EIDs of an IBA Proteus ONE proton therapy system. The PCD-CT scans followed the same acquisition protocol as the clinical EID-CBCT for head imaging. Geometric distortions were estimated and corrected during image reconstruction. To assess nonspectral image quality, standardized phantoms with known material compositions were scanned under matched x-ray exposure and beam width conditions. The resulting PCD-CT images were compared with EID-CBCT in terms of modulation transfer function (MTF) and contrast-to-noise ratio (CNR), defined based on 15 different material inserts from 2 physical phantoms. In addition, spectral PCD-CT data were used to estimate the electron density ratio of each material relative to water. The estimated values were then compared with reference values. RESULTS: Compared with the EID, the PCD exhibits significantly higher detective quantum efficiency, especially at high spatial frequencies. When the MTFs of PCD-CT and EID-CBCT were matched by adjusting the reconstruction kernel, PCD-CT consistently achieved higher CNR across all material inserts tested. For electron density ratios estimated from spectral PCD-CT, the mean absolute percent error was 1.8%. CONCLUSIONS: The feasibility of gantry-mounted PCD tomographic imaging for proton therapy has been demonstrated for the first time. The PCD-CT prototype provides superior nonspectral image quality compared with conventional EID-CBCT. Additionally, its spectral imaging capability enables accurate estimation of electron density ratios.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin.
- Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas; Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas.
- Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas.
- Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas; Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas. Electronic address: Hao.Gao2@…
- Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin; Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas; Department of Interventional Radiology, U…
Förderung
- NIBIB NIH HHS
Zitieren
APA
De Caro, C., Lin, Y., Setianegara, J., Chen, G. H., Gao, H., & Li, K. (2026). A Gantry-Mounted Photon-Counting Detector Computed Tomography Prototype for Image Guided Proton Therapy. International Journal of Radiation Oncology*Biology*Physics. https://doi.org/10.1016/j.ijrobp.2026.02.210
Vancouver
De Caro C, Lin Y, Setianegara J, Chen GH, Gao H, Li K. A Gantry-Mounted Photon-Counting Detector Computed Tomography Prototype for Image Guided Proton Therapy. International Journal of Radiation Oncology*Biology*Physics. 2026. doi:10.1016/j.ijrobp.2026.02.210
BibTeX
@article{DeCaro2026,
author = {De Caro, Christian and Lin, Yuting and Setianegara, Jufri and Chen, Guang-Hong and Gao, Hao and Li, Ke},
title = {A Gantry-Mounted Photon-Counting Detector Computed Tomography Prototype for Image Guided Proton Therapy},
journal = {International Journal of Radiation Oncology*Biology*Physics},
year = {2026},
doi = {10.1016/j.ijrobp.2026.02.210}
}
EUVIMED