Towards a more realistic anthropomorphic chest phantom using 3D‐printed and cork‐integrated components

Joost F Hop, Ivan Dudurych, Thom R G Stams, Geertruida H de Bock, Rozemarijn Vliegenthart, Marcel J W Greuter

Medical Physics · 2025

DOI: 10.1002/mp.17956 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Joost F Hop and 5 co-authors. Institution: Department of Radiology, University Medical Center Groningen, Groningen, the Netherlands. Major MEDLINE topics: Printing, Three-Dimensional, Phantoms, Imaging, Tomography, X-Ray Computed, Thorax. Cited 3 times so far (Crossref); 41 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Thorax phantoms for computed tomography (CT) imaging often lack realistic lung parenchyma and bronchovascular anatomy. To improve anatomical accuracy, 3D‐printed chest phantoms have been developed as more realistic alternatives to existing models.

Purpose

To evaluate whether an in‐house developed anthropomorphic phantom insert realistically represents the anatomical structures and attenuation characteristics compared to the original phantom and human CT data.

Methods

The anthropomorphic chest phantom “Lungman” was modified by integrating a 3D‐printed insert, cork‐based lung parenchyma, and lung nodules. The phantom was scanned on a CT system and evaluated using qualitative and quantitative CT analyses, comparing attenuation values and histogram distributions to human CT data. Subjective radiologist assessments were conducted to compare anatomical realism between the modified and unmodified phantom.

Results

Qualitative assessment of CT value distribution showed strong similarity between the modified phantom and human lung parenchyma, although the radiodensity characteristics of the 3D‐printed bronchovascular insert still require further refinement. Quantitative analysis confirmed that the modified phantom's parenchymal attenuation (−854 Hounsfield unit [HU]) closely matched human lung parenchyma (−872 HU, p > 0.05), whereas the unmodified phantom showed lower attenuation (−997 HU, p < 0.05). However, the bronchovascular insert showed lower attenuation than human vasculature (−41 HU vs. 42 HU, p < 0.05). Radiologists rated the modified phantom higher for anatomical realism.

Conclusions

This study presents a modified, anthropomorphic thorax phantom that integrates realistic 3D‐printed bronchial and vascular structures, lung parenchyma, and lung nodules.

Subjects

MeSH terms (MEDLINE)

Printing, Three-DimensionalPhantoms, ImagingTomography, X-Ray ComputedThoraxRadiography, ThoracicHumansLung

Author keywords

3D‐printinganthropomorphic thorax phantomcomputed tomographylung cancerpulmonary nodulesquality assurance

Authors and institutions

  • Department of Radiology, University Medical Center Groningen, Groningen, the Netherlands.
  • Department of Epidemiology, University Medical Center Groningen, Groningen, the Netherlands.

Cite

APA

Hop, J. F., Dudurych, I., Stams, T. R. G., de Bock, G. H., Vliegenthart, R., & Greuter, M. J. W. (2025). Towards a more realistic anthropomorphic chest phantom using 3D‐printed and cork‐integrated components. Medical Physics. https://doi.org/10.1002/mp.17956

Vancouver

Hop JF, Dudurych I, Stams TRG, de Bock GH, Vliegenthart R, Greuter MJW. Towards a more realistic anthropomorphic chest phantom using 3D‐printed and cork‐integrated components. Medical Physics. 2025. doi:10.1002/mp.17956

BibTeX
@article{Hop2025,
  author = {Hop, Joost F and Dudurych, Ivan and Stams, Thom R G and de Bock, Geertruida H and Vliegenthart, Rozemarijn and Greuter, Marcel J W},
  title = {Towards a more realistic anthropomorphic chest phantom using 3D‐printed and cork‐integrated components},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17956}
}
Start for free →