EUVIMEDSkopri EUVIMED
Open access · CC BY 4.0

Model‐based perfusion reconstruction with time separation technique in cone‐beam CT dynamic liver perfusion imaging

Hana Haseljić, Robert Frysch, Vojtěch Kulvait, Thomas Werncke, Inga Brüsch, Oliver Speck, Jessica Schulz, Michael Manhart, Georg Rose

Medical Physics · 2025

DOI: 10.1002/mp.17652 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Hana Haseljić and 8 co-authors. Institution: Institute for Medical Engineering, Otto von Guericke University Magdeburg, Magdeburg, Germany. Major MEDLINE topics: Cone-Beam Computed Tomography, Liver, Perfusion Imaging, Image Processing, Computer-Assisted. Cited 1 times so far (Crossref); 39 references. The full text is freely available (open access, CC BY 4.0). Funded by: Bundesministerium für Bildung und Forschung.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

The success of embolization, a minimally invasive treatment of liver cancer, could be evaluated in the operational room with cone‐beam CT by acquiring a dynamic perfusion scan to inspect the contrast agent flow.

Purpose

The reconstruction algorithm must address the issues of low temporal sampling and higher noise levels inherent in cone‐beam CT systems, compared to conventional CT.

Methods

Therefore, a model‐based perfusion reconstruction based on the time separation technique (TST) was applied. TST uses basis functions to model time attenuation curves. These functions are either analytical or based on prior knowledge (PK), extracted using singular value decomposition of the classical CT perfusion data of animal subjects. To explore how well the PK can model perfusion dynamics and what the potential limitations are, the dynamic cone‐beam CT (CBCT) perfusion scan was simulated from a dynamic CT perfusion scan under different noise levels. The TST method was compared to static reconstruction.

Results

It was demonstrated on this simulated dynamic CBCT perfusion scan that a set consisting of only four basis functions results in perfusion maps that preserve relevant information, denoise the data, and outperform static reconstruction under higher noise levels. TST with PK would not only outperform static reconstruction but also the TST with analytical basis functions. Furthermore, it has been shown that only eight CBCT rotations, unlike previously assumed ten, are sufficient to obtain the perfusion maps comparable to the reference CT perfusion maps. This contributes to saving dose and reconstruction time. The real dynamic CBCT perfusion scan, reconstructed under the same conditions as the simulated scan, shows potential for maintaining the accuracy of the perfusion maps. By visual inspection, the embolized region was matching to that in corresponding CT perfusion maps.

Conclusions

CBCT reconstruction of perfusion scan data using the TST method has shown promising potential, outperforming static reconstructions and potentially saving dose by reducing the necessary number of acquisition sweeps. Further analysis of a larger cohort of patient data is needed to draw final conclusions regarding the expected advantages of the TST.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyLiverPerfusion ImagingImage Processing, Computer-AssistedAnimalsTime FactorsAlgorithmsSignal-To-Noise Ratio

Author keywords

cone‐beam CT dynamic liver perfusion imagingprior knowledgetime separation technique

Authors and institutions

  • Institute for Medical Engineering, Otto von Guericke University Magdeburg, Magdeburg, Germany.
  • Research Campus STIMULATE, Otto von Guericke University Magdeburg, Magdeburg, Germany.
  • Institute of Materials Physics, Helmholtz-Zentrum Hereon, Geesthacht, Germany.
  • Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Institute for Laboratory Animal Science, Hannover Medical School, Hannover, Germany.
  • Siemens Healthineers AG, Forchheim, Germany.

Funding

  • Bundesministerium für Bildung und Forschung

Cite

APA

Haseljić, H., Frysch, R., Kulvait, V., Werncke, T., Brüsch, I., Speck, O., Schulz, J., Manhart, M., & Rose, G. (2025). Model‐based perfusion reconstruction with time separation technique in cone‐beam CT dynamic liver perfusion imaging. Medical Physics. https://doi.org/10.1002/mp.17652

Vancouver

Haseljić H, Frysch R, Kulvait V, Werncke T, Brüsch I, Speck O, et al. Model‐based perfusion reconstruction with time separation technique in cone‐beam CT dynamic liver perfusion imaging. Medical Physics. 2025. doi:10.1002/mp.17652

BibTeX
@article{Haselji2025,
  author = {Haseljić, Hana and Frysch, Robert and Kulvait, Vojtěch and Werncke, Thomas and Brüsch, Inga and Speck, Oliver and Schulz, Jessica and Manhart, Michael and Rose, Georg},
  title = {Model‐based perfusion reconstruction with time separation technique in cone‐beam CT dynamic liver perfusion imaging},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17652}
}
Start for free →