Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT
PLOS One · 2026
DOI: 10.1371/journal.pone.0349360 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, 2026. Verfasst von Falk L Wiegmann und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, The University of British Columbia, Vancouver, British Columbia, Canada. Hauptthemen laut MEDLINE: Algorithms, X-Ray Microtomography, Cone-Beam Computed Tomography, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: British Columbia Lung Association.
Abstract (Originalsprache)
The ramp filter kernel and cutoff frequency are fundamental parameters of the Feldkamp–Davis–Kress (FDK) algorithm that determine the resolution and noise characteristics of the reconstructed image. Despite their importance, the filter configuration is frequently unreported in preclinical micro-CT studies. We reconstruct identical data from a GE eXplore CT 120 scanner using four filter kernels (ramp, Shepp-Logan, cosine, Hamming) at four cutoff frequencies (1.0, 0.8, 0.6, and 0.379 × Nyquist, matched to the detector-to-voxel size ratio) and evaluate each of the sixteen configurations using the modulation transfer function (MTF), noise power spectrum (NPS), and non-prewhitening detectability index (NPW d ′ ). Qualitative assessment is performed on a mouse lung specimen. Across the sixteen configurations, MTF 10 ranges from 1.74 to 3.22 lp/mm, integrated NPS from 75 670 to 13 259 HU 2 , and the Rose criterion crossing diameter from 2.03 to 0.90 mm at Δ C = 500 H U and from 6.89 to 3.56 mm at 100 HU. This note presents the data as a concise, primarily comparative visual and quantitative reference for FDK filter selection in preclinical cone-beam CT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Physics and Astronomy, The University of British Columbia, Vancouver, British Columbia, Canada.
- Department of Oral Biological and Medical Sciences, The University of British Columbia, Vancouver, British Columbia, Canada.
ORCID-Profile: Falk L Wiegmann
Förderung
- British Columbia Lung Association
Zitieren
APA
Wiegmann, F. L., & Ford, N. L. (2026). Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT. PLOS One. https://doi.org/10.1371/journal.pone.0349360
Vancouver
Wiegmann FL, Ford NL. Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT. PLOS One. 2026. doi:10.1371/journal.pone.0349360
BibTeX
@article{Wiegmann2026,
author = {Wiegmann, Falk L and Ford, Nancy L},
title = {Resolution–noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT},
journal = {PLOS One},
year = {2026},
doi = {10.1371/journal.pone.0349360}
}
EUVIMED