A magnetic field compatible readout circuit for enhanced coincidence time resolution in BGO Cherenkov radiation–based TOF‐PET detectors
Medical Physics · 2025
DOI: 10.1002/mp.17643 ↗At a glance
Journal article in Medical Physics, 2025. Written by Shirin Pourashraf and 2 co-authors. Institution: Department of Radiology, Stanford University, Stanford, California, USA. Major MEDLINE topics: Positron-Emission Tomography, Magnetic Fields, Bismuth, Germanium. Cited 2 times so far (Crossref); 13 references. Funded by: Stanford Cancer Imaging Training, NIH HHS, National Institutes of Health.
Abstract (original language)
Background
Developing time‐of‐flight positron emission tomography/magnetic resonance imaging (TOF‐PET/MRI) detectors that exploit prompt Cherenkov photons from bismuth germanate (BGO) crystals for estimating 511 keV photon arrival time.
Purpose
To present a low‐noise, high‐speed electronic readout circuit design for BGO‐based TOF‐PET detectors that achieves enhanced coincidence time resolution (CTR) in presence of a strong magnetic field.
Methods
The CTR of a BGO‐based TOF‐PET test detector employing a high‐speed, low‐noise electronic readout chain was evaluated in a strong magnetic field produced by a permanent magnet placed directly on top of the circuit. For these experiments, which exploit Cherenkov radiation for precise measurement of annihilation photon time arrival time difference, a point source of 22 Na was positioned between a pair of 3 × 3 × 15 mm 3 polished BGO crystals wrapped in Teflon tape and optically coupled to 3 × 3 mm 2 ultra‐violet (UV)‐sensitive silicon photomultipliers (SiPMs).
Results
By incorporating both Cherenkov (prompt) and standard (slow) luminescence components, 283 ± 8 ps and 275 ± 10 ps full‐width‐half‐maximum (FWHM) CTR were achieved without and with the permanent magnet present, respectfully. These values improved to 236 ± 4 ps and 216 ± 17 ps FWHM when only the Cherenkov components of the timing signal (events with the fastest rise time) were considered.
Conclusions
Results indicate we have designed a high‐performance readout circuit that achieves significantly the same CTR in BGO with or without a strong magnetic field present. This further demonstrates that UV SiPMs can effectively operate in a strong magnetic field while remaining highly advantageous for detecting Cherenkov radiation, thus highlighting their potential to be used in BGO‐based TOF‐PET/MRI scanners.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Radiology, Stanford University, Stanford, California, USA.
- Applied Nuclear Physics Program, Lawrence Berkeley National Laboratory, Berkeley, California, USA.
- Molecular Imaging Program, Departments of Radiology, Bioengineering, Physics and Electrical Engineering, Stanford University, Stanford, California, USA.
Funding
- Stanford Cancer Imaging Training
- NIH HHS
- National Institutes of Health
Cite
APA
Pourashraf, S., Cates, J. W., & Levin, C. S. (2025). A magnetic field compatible readout circuit for enhanced coincidence time resolution in BGO Cherenkov radiation–based TOF‐PET detectors. Medical Physics. https://doi.org/10.1002/mp.17643
Vancouver
Pourashraf S, Cates JW, Levin CS. A magnetic field compatible readout circuit for enhanced coincidence time resolution in BGO Cherenkov radiation–based TOF‐PET detectors. Medical Physics. 2025. doi:10.1002/mp.17643
BibTeX
@article{Pourashraf2025,
author = {Pourashraf, Shirin and Cates, Joshua W and Levin, Craig S},
title = {A magnetic field compatible readout circuit for enhanced coincidence time resolution in BGO Cherenkov radiation–based TOF‐PET detectors},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17643}
}
EUVIMED