Design and characterization of a hybrid PET detector with DOI capability

Wen He, Yangyang Zhao, Honghao Zeng, Wenjie Huang, Hang Yang, Xin Zhao, Qiang Wang, Lu Wang, Ming Niu, Lei Zhang, Qiushi Ren, Zheng Gu

Medical Physics · 2024

DOI: 10.1002/mp.17313 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Wen He und 11 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Biomedical Engineering, Shenzhen Bay Laboratory, Shenzhen, China. Hauptthemen laut MEDLINE: Positron-Emission Tomography, Equipment Design. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 40 Literaturangaben. Gefördert von: National Natural Science Foundation of China, GuangDong Basic and Applied Basic Research Foundation, National Key R&D Program of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Monolithic or semi‐monolithic detectors are attractive for positron emission tomography (PET) scanners with depth‐of‐interaction (DOI) capability. However, they often require complicated calibrations to determine the interaction positions of gamma photons.

Purpose

We introduce a novel hybrid detector design that combines pixelated and semi‐monolithic elements to achieve DOI capability while simplifying the calibrations for positioning.

Methods

A prototype detector with eight hybrid lutetium–yttrium oxyorthosilicate (LYSO) layers having dimensions of 25.8 × 12.9 × 15 mm 3 was constructed. The energy‐weighted and energy‐squared weighted averages were used for estimating the x ‐ (pixelated direction) and y ‐positions (non‐pixelated direction). Pseudo‐pixels were defined as discrete areas on the flood image based on the crystal look‐up table (LUT). The intrinsic spatial resolutions in the pixelated and non‐pixelated directions were measured. The ratio of the maximum to the sum of the multipixel photon counter (MPPC) signals was used to estimate the DOI positions. The coincidence timing resolution (CTR) was measured using the average and energy‐weighted average of the earliest n time stamps. Two energy windows of 250–700 and 400–600 keV were applied for the measurements.

Results

The pattern of the flood images showed discrete event clusters, demonstrating that simple calibrations for determining the x‐ and y ‐positions of events could be achieved. Under 400–600 keV energy window, the average intrinsic spatial resolutions were 1.15 and 1.34 mm for the pixelated and non‐pixelated directions; the average DOI resolution of the second row of pseudo‐pixels was 5.1 mm in full width at half maximum (FWHM); when using the energy‐weighted average of the earliest four‐time stamps, the best CTR of 350 ps was achieved. Applying a broader energy window of 250–700 keV only slightly degrades the DOI resolution while maintaining the intrinsic resolution; the best CTR degrades to 410 ps.

Conclusions

The proposed hybrid detector concept was verified, and a prototype detector showed high performance for 3D positioning and timing resolution. The novel detector concept shows promise for preclinical and clinical PET scanners with DOI capability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Positron-Emission TomographyEquipment DesignCalibrationPhotons

Stichwörter der Autor:innen

calibrationcoincidence timing resolution (CTR)depth‐of‐interaction (DOI)positron emission tomography (PET)semi‐monolithicspatial resolution

Autor:innen und Institutionen

  • Institute of Biomedical Engineering, Shenzhen Bay Laboratory, Shenzhen, China.
  • Peking University Shenzhen Graduate School, Shenzhen, China.
  • Ciechanover Institute of Precision and Regenerative Medicine, School of Medicine, The Chinese University of Hong Kong, Shenzhen, China.
  • 26th Institute of China Electronics Technology Group Corporation, Chongqing, China.
  • Institute of Cancer Research, Shenzhen Bay Laboratory, Shenzhen, China.

Förderung

  • National Natural Science Foundation of China
  • GuangDong Basic and Applied Basic Research Foundation
  • National Key R&D Program of China

Zitieren

APA

He, W., Zhao, Y., Zeng, H., Huang, W., Yang, H., Zhao, X., Wang, Q., Wang, L., Niu, M., Zhang, L., Ren, Q., & Gu, Z. (2024). Design and characterization of a hybrid PET detector with DOI capability. Medical Physics. https://doi.org/10.1002/mp.17313

Vancouver

He W, Zhao Y, Zeng H, Huang W, Yang H, Zhao X, et al. Design and characterization of a hybrid PET detector with DOI capability. Medical Physics. 2024. doi:10.1002/mp.17313

BibTeX
@article{He2024,
  author = {He, Wen and Zhao, Yangyang and Zeng, Honghao and Huang, Wenjie and Yang, Hang and Zhao, Xin and Wang, Qiang and Wang, Lu and Niu, Ming and Zhang, Lei and Ren, Qiushi and Gu, Zheng},
  title = {Design and characterization of a hybrid PET detector with DOI capability},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17313}
}