On the origin of MTF reduction in grating‐based x‐ray differential phase contrast CT imaging
Medical Physics · 2025
DOI: 10.1002/mp.17593 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Yuhang Tan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences,…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Guangdong Basic and Applied Basic Research Foundation, Youth Innovation Promotion Association of the Chinese Academy of Sciences.
Abstract (Originalsprache)
Background
The complementary absorption contrast CT (ACT) and differential phase contrast CT (DPCT) can be generated simultaneously from an x‐ray computed tomography (CT) imaging system incorporated with grating interferometer. However, it has been reported that ACT images exhibit better spatial resolution than DPCT images. By far, the primary cause of such discrepancy remains unclear.
Purpose
The purpose of this study is to investigate the underlying cause of the resolution discrepancy between ACT and DPCT in a grating interferometer CT imaging system.
Methods
In this study, theoretical derivations were performed with a ‐phase Talbot–Lau grating interferometer system to model the signal formation mechanism of absorption imaging and phase imaging, respectively. In addition, physical, and numerical experiments were conducted to verify the theoretical findings and assess the resolution discrepancy between ACT and DPCT under various conditions. Herein, the ACT and DPCT images were reconstructed from the filtered‐back‐projection algorithm using a standard Ramp filter and a standard Hilbert filter, respectively.
Results
Experiments demonstrated that the spatial resolution of ACT and DPCT images are primarily impacted by the beam diffraction induced signal splitting. In particular, lower modulation transfer function (MTF) was observed for DPCT than ACT due to the opposite‐superposition of phase signals. In addition, factors such as focal spot size, beam spectra, object composition, sample size, and detector pixel size were found to have minor impacts on the MTFs of both ACT and DPCT.
Conclusions
In conclusion, this study reveals that the opposite‐superposition of split phase signals causes the spatial resolution reduction in DPCT imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
- Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
- Key Laboratory of Biomedical Imaging Science and System, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
- National Innovation Center for Advanced Medical Devices, Shenzhen, Guangdong, China.
- Institute of High Energy Physics, Chinese Academy of Sciences, Beijing, China.
Förderung
- National Natural Science Foundation of China
- Guangdong Basic and Applied Basic Research Foundation
- Youth Innovation Promotion Association of the Chinese Academy of Sciences
Zitieren
APA
Tan, Y., Yang, J., Zheng, H., Liang, D., Zhu, P., & Ge, Y. (2025). On the origin of MTF reduction in grating‐based x‐ray differential phase contrast CT imaging. Medical Physics. https://doi.org/10.1002/mp.17593
Vancouver
Tan Y, Yang J, Zheng H, Liang D, Zhu P, Ge Y. On the origin of MTF reduction in grating‐based x‐ray differential phase contrast CT imaging. Medical Physics. 2025. doi:10.1002/mp.17593
BibTeX
@article{Tan2025,
author = {Tan, Yuhang and Yang, Jiecheng and Zheng, Hairong and Liang, Dong and Zhu, Peiping and Ge, Yongshuai},
title = {On the origin of MTF reduction in grating‐based x‐ray differential phase contrast CT imaging},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17593}
}
EUVIMED