Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study

Lingxuan Leng iD, Qizhen Zhu, HuiYing Qu, Bing Zhou, Wenbo Li, Fuquan Zhang, Bo Yang, Jie Qiu

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70423 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Lingxuan Leng und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijin…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Image Processing, Computer-Assisted, Algorithms. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key R&D Program of China, Ministry of Science and Technology of the People's Republic of China, National Key Research and Development Program of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aims to systematically evaluate the impact of different CT acquisition and reconstruction parameters on the accuracy of dual‐layer spectral CT‐based functional imaging, including effective atomic number (Z eff ) maps, relative electron density (RED) maps, and virtual monochromatic images (VMIs).

Methods

A standardized phantom equipped with various material inserts, including air, soft‐tissue–equivalent, polymer, and bone‐equivalent materials, was scanned using a dual‐layer spectral CT system. The investigated acquisition and reconstruction parameters included tube voltage, tube current–time product (mAs), reconstruction thickness, pitch, and reconstruction filter. Quantitative accuracy was assessed for RED and Z eff by comparison with manufacturer‐provided reference values, using Bland–Altman analysis. Comparative analyses were performed between VMIs at 40, 70, and 100 keV and conventional CT (120 kVp single‐energy reconstruction from the dual‐layer system) by calculating the mean and standard deviation of Hounsfield units (HU) differences, to evaluate the dependence of CT number stability on acquisition and reconstruction parameters.

Results

Across all acquisition and reconstruction parameters, RED and Z eff measurements remained accurate, with signed differences within 2.4% and 0.4%, respectively. Compared with conventional CT, VMIs demonstrated greater robustness to parameter change. At 70 keV, HU variations did not exceed 8 HU, whereas conventional CT exhibited deviations up to 118 HU. Low‐energy VMIs (40 keV) showed greater variability, with a maximum difference of 27 HU. Among the investigated parameters, tube voltage exerted the most evident influence.

Conclusion

Dual‐layer spectral CT demonstrated quantitative accuracy in functional imaging (ED, Z eff , and VMIs) across diverse acquisition and reconstruction parameters.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, X-Ray ComputedImage Processing, Computer-AssistedAlgorithmsRadiographic Image Interpretation, Computer-AssistedRadiotherapy Planning, Computer-AssistedHumans

Stichwörter der Autor:innen

effective atomic numberradiation therapy planningrelative electron densityspectral CTvirtual monochromatic images

Autor:innen und Institutionen

  • Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • Department of Radiation Oncology, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.

ORCID-Profile: Lingxuan Leng

Förderung

  • National Key R&D Program of China
  • Ministry of Science and Technology of the People's Republic of China
  • National Key Research and Development Program of China

Zitieren

APA

Leng, L., Zhu, Q., Qu, H., Zhou, B., Li, W., Zhang, F., Yang, B., & Qiu, J. (2025). Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70423

Vancouver

Leng L, Zhu Q, Qu H, Zhou B, Li W, Zhang F, et al. Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70423

BibTeX
@article{Leng2025,
  author = {Leng, Lingxuan and Zhu, Qizhen and Qu, HuiYing and Zhou, Bing and Li, Wenbo and Zhang, Fuquan and Yang, Bo and Qiu, Jie},
  title = {Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70423}
}
Kostenlos starten →