Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70423 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Lingxuan Leng und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijin…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Image Processing, Computer-Assisted, Algorithms. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key R&D Program of China, Ministry of Science and Technology of the People's Republic of China, National Key Research and Development Program of China.
Abstract (Originalsprache)
Purpose
This study aims to systematically evaluate the impact of different CT acquisition and reconstruction parameters on the accuracy of dual‐layer spectral CT‐based functional imaging, including effective atomic number (Z eff ) maps, relative electron density (RED) maps, and virtual monochromatic images (VMIs).
Methods
A standardized phantom equipped with various material inserts, including air, soft‐tissue–equivalent, polymer, and bone‐equivalent materials, was scanned using a dual‐layer spectral CT system. The investigated acquisition and reconstruction parameters included tube voltage, tube current–time product (mAs), reconstruction thickness, pitch, and reconstruction filter. Quantitative accuracy was assessed for RED and Z eff by comparison with manufacturer‐provided reference values, using Bland–Altman analysis. Comparative analyses were performed between VMIs at 40, 70, and 100 keV and conventional CT (120 kVp single‐energy reconstruction from the dual‐layer system) by calculating the mean and standard deviation of Hounsfield units (HU) differences, to evaluate the dependence of CT number stability on acquisition and reconstruction parameters.
Results
Across all acquisition and reconstruction parameters, RED and Z eff measurements remained accurate, with signed differences within 2.4% and 0.4%, respectively. Compared with conventional CT, VMIs demonstrated greater robustness to parameter change. At 70 keV, HU variations did not exceed 8 HU, whereas conventional CT exhibited deviations up to 118 HU. Low‐energy VMIs (40 keV) showed greater variability, with a maximum difference of 27 HU. Among the investigated parameters, tube voltage exerted the most evident influence.
Conclusion
Dual‐layer spectral CT demonstrated quantitative accuracy in functional imaging (ED, Z eff , and VMIs) across diverse acquisition and reconstruction parameters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
- Department of Radiation Oncology, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.
ORCID-Profile: Lingxuan Leng
Förderung
- National Key R&D Program of China
- Ministry of Science and Technology of the People's Republic of China
- National Key Research and Development Program of China
Zitieren
APA
Leng, L., Zhu, Q., Qu, H., Zhou, B., Li, W., Zhang, F., Yang, B., & Qiu, J. (2025). Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70423
Vancouver
Leng L, Zhu Q, Qu H, Zhou B, Li W, Zhang F, et al. Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70423
BibTeX
@article{Leng2025,
author = {Leng, Lingxuan and Zhu, Qizhen and Qu, HuiYing and Zhou, Bing and Li, Wenbo and Zhang, Fuquan and Yang, Bo and Qiu, Jie},
title = {Impact of acquisition and reconstruction parameters on quantitative accuracy in dual‐layer spectral CT: A phantom study},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70423}
}
EUVIMED