On the acceptance, commissioning, and quality assurance of electron FLASH units
Medical Physics · 2025-02 · 52 (2) · 1207-1223
DOI: 10.1002/mp.17483 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 27.10.2024. Verfasst von Allison Palmiero und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, James Cancer Hospital and Solove Research Institute, The Ohio State University, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Electrons, Particle Accelerators, Quality Assurance, Health Care. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 64 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, MD Anderson's Division of Radiation Oncology, Cancer Prevention and Research Institute of Texas.
Abstract (Originalsprache)
Background and purpose FLASH or ultra‐high dose rate (UHDR) radiation therapy (RT) has gained attention in recent years for its ability to spare normal tissues relative to conventional dose rate (CDR) RT in various preclinical trials. However, clinical implementation of this promising treatment option has been limited because of the lack of availability of accelerators capable of delivering UHDR RT. Commercial options are finally reaching the market that produce electron beams with average dose rates of up to 1000 Gy/s. We established a framework for the acceptance, commissioning, and periodic quality assurance (QA) of electron FLASH units and present an example of commissioning.
Methods
A protocol for acceptance, commissioning, and QA of UHDR linear accelerators was established by combining and adapting standards and professional recommendations for standard linear accelerators based on the experience with UHDR at four clinical centers that use different UHDR devices. Non‐standard dosimetric beam parameters considered included pulse width, pulse repetition frequency, dose per pulse, and instantaneous dose rate, together with recommendations on how to acquire these measurements.
Results
The 6‐ and 9‐MeV beams of an UHDR electron device were commissioned by using this developed protocol. Measurements were acquired with a combination of ion chambers, beam current transformers (BCTs), and dose‐rate–independent passive dosimeters. The unit was calibrated according to the concept of redundant dosimetry using a reference setup.
Conclusion
This study provides detailed recommendations for the acceptance testing, commissioning, and routine QA of low‐energy electron UHDR linear accelerators. The proposed framework is not limited to any specific unit, making it applicable to all existing eFLASH units in the market. Through practical insights and theoretical discourse, this document establishes a benchmark for the commissioning of UHDR devices for clinical use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, James Cancer Hospital and Solove Research Institute, The Ohio State University, Columbus, Ohio, USA.
- Division of Radiation Oncology, Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Graduate School of Biomedical Sciences, The University of Texas, Houston, Texas, USA.
- INSERM U1030, Gustave Roussy, Université Paris-Saclay, Villejuif, France.
- Division of Radiation Oncology, Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Institute of Radiation Physics, Lausanne University Hospital and Lausanne University, Lausanne, Switzerland.
Förderung
- NCI NIH HHS
- MD Anderson's Division of Radiation Oncology
- Cancer Prevention and Research Institute of Texas
- UTHealth Innovation for Cancer Prevention Research Training Program Pre-doctoral Fellowship
- National Cancer Institute of the National Institutes of Health Cancer Center Support Grant
- University Cancer Foundation via the Institutional Research Grant program at MD Anderson Cancer Center
- National Cancer Institute of the National Institutes of Health
Zitieren
APA
Palmiero, A., Liu, K., Colnot, J., Chopra, N., Neill, D., Connell, L., Velasquez, B., Koong, A. C., Lin, S. H., Balter, P., Tailor, R., Robert, C., Germond, J. F., Gonçalves Jorge, P., Geyer, R., Beddar, S., Moeckli, R., & Schüler, E. (2025). On the acceptance, commissioning, and quality assurance of electron FLASH units. Medical Physics, 52(2), 1207-1223. https://doi.org/10.1002/mp.17483
Vancouver
Palmiero A, Liu K, Colnot J, Chopra N, Neill D, Connell L, et al. On the acceptance, commissioning, and quality assurance of electron FLASH units. Medical Physics. 2025;52(2):1207-1223. doi:10.1002/mp.17483
BibTeX
@article{Palmiero2025,
author = {Palmiero, Allison and Liu, Kevin and Colnot, Julie and Chopra, Nitish and Neill, Denae and Connell, Luke and Velasquez, Brett and Koong, Albert C and Lin, Steven H and Balter, Peter and Tailor, Ramesh and Robert, Charlotte and Germond, Jean-François and Gonçalves Jorge, Patrik and Geyer, Reiner and Beddar, Sam and Moeckli, Raphael and Schüler, Emil},
title = {On the acceptance, commissioning, and quality assurance of electron FLASH units},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {2},
pages = {1207--1223},
doi = {10.1002/mp.17483}
}
EUVIMED