Evaluation of a Faraday cup‐style detector as a beam diagnostic system for ultra‐high dose rate (FLASH) electron beams

Alan Lopez, Alexander Baikalov, Kevin Liu iD, Nolan Esplen, Stefan Bartzsch, Emil Schüler

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70819 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Alan Lopez und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Electrons, Particle Accelerators, Phantoms, Imaging, Radiometry. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, MD Anderson's Division of Radiation Oncology; by Cancer Center Support (Core), NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Beam diagnostic systems are essential for the development and quality assurance of ultra‐high dose rate (UHDR) radiation sources to enable reliable delivery of FLASH radiotherapy (RT). Critically, suitable beam diagnostic systems for electron FLASH sources are lacking.

Purpose

In this study, we evaluated a Faraday cup–style beam collector (BC), the BC‐145‐Al, on an UHDR electron linac for FLASH‐RT applications.

Methods

The BC performance was benchmarked against a reference alternating‐current current transformer (ACCT) across a wide range of beam currents and pulse structures to test the BC's charge linearity and signal resolution capabilities. Two termination impedances were tested: 200 kΩ and 13 Ω.

Results

The highly time‐resolved signal of the BC was charge‐proportional when terminated at 200 kΩ and current‐proportional when terminated at 13 Ω, agreeing well with the time‐resolved ACCT signal. The cumulative signals from the BC and ACCT showed excellent agreement, maintaining linearity with a deviation of less than ± 0.5% over 50 consecutive pulses (4‐µs pulse width, 120‐Hz pulse repetition frequency [PRF]). The BC/ACCT signal ratio remained stable within ± 0.5% as the PRF increased from 5 to 120 Hz at constant pulse width. Pulse width measurements from both detectors were also consistent within ± 0.5% across all nominal pulse width values (0.5–4 µs). When the charge per pulse was modulated by either increasing the pulse width or decreasing the source‐to‐surface distance (from 32.3 to 19.3 cm), the BC signal remained approximately linear with respect to the ACCT signal; however, the BC/ACCT ratio deviated by up to ± 6% across this range.

Conclusions

The BC produced highly time‐resolved absolute charge measurements of a pulsed UHDR electron beam across a large range of beam parameters. These results support its potential as a beam diagnostic system for both preclinical studies and clinical applications of electron FLASH radiotherapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsParticle AcceleratorsPhantoms, ImagingRadiometryRadiotherapy Planning, Computer-AssistedQuality Assurance, Health CareHumansRadiotherapy DosageEquipment Design

Stichwörter der Autor:innen

FLASH radiotherapyFaraday cupbeam diagnosticsultra‐high dose rate

Autor:innen und Institutionen

  • Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • The University of Texas MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, USA.
  • Department of Radiation Oncology, School of Medicine and Klinikum rechts der Isar, Technical University of Munich, Munich, Germany.
  • Institute of Radiation Medicine, Helmholtz Zentrum München GmbH, German Research Center for Environmental Health, Neuherberg, Germany.

ORCID-Profile: Kevin Liu

Förderung

  • NCI NIH HHS
  • MD Anderson's Division of Radiation Oncology; by Cancer Center Support (Core)
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Lopez, A., Baikalov, A., Liu, K., Esplen, N., Bartzsch, S., & Schüler, E. (2026). Evaluation of a Faraday cup‐style detector as a beam diagnostic system for ultra‐high dose rate (FLASH) electron beams. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70819

Vancouver

Lopez A, Baikalov A, Liu K, Esplen N, Bartzsch S, Schüler E. Evaluation of a Faraday cup‐style detector as a beam diagnostic system for ultra‐high dose rate (FLASH) electron beams. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70819

BibTeX
@article{Lopez2026,
  author = {Lopez, Alan and Baikalov, Alexander and Liu, Kevin and Esplen, Nolan and Bartzsch, Stefan and Schüler, Emil},
  title = {Evaluation of a Faraday cup‐style detector as a beam diagnostic system for ultra‐high dose rate (FLASH) electron beams},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70819}
}
Kostenlos starten →