A slope method for the determination of electron energy for quality assurance
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14369 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Larry W Hill und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Therapy, Genesis Care/21st Century Oncology, Fort Walton Beach, Florida, USA. Hauptthemen laut MEDLINE: Electrons, Quality Assurance, Health Care, Particle Accelerators. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Particle accelerators, manufactured for delivering patient radiation treatment, require numerous and frequent quality assurance measures. One of those is the periodic check for electron energy stability. The American Association of Physicists in Medicine has established requirements for this procedure. The current recommendation is to perform a ratio of two ionization points, one at D max and another at a point approximately to the 50% depth, compared to a baseline as a relative check.
Purpose
This ratio method is a sensitive measurement and sometimes produces results that are difficult to interpret or relate to acceptable tolerances. We sought to find a simple method that gives more stable results, which can be interpreted and related to energy changes.
Method
We propose a method that takes two measurements on the descending portion of the shifted percent depth ionization (PDI) curves to calculate the slope, tangent to the I 50 point, the point at which the ionization falls to 50% of its maximum value. We then used the slope measurement, compared to an established baseline, to relate energy.
Results
After collecting data over a 3‐year period, we saw that standard deviations for the slope method have much less variability than the traditional ratio method. We were also able to correlate the slope results to ionization scans performed in water and found they were in better agreement than the traditional ratio method.
Conclusion
The slope method does not require precise positioning since the slope remains relatively constant over the descending portion of the curve. Our data show that this results in an easier interpretative test of electron energy stability and delivers reliable feedback for quality assurance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Therapy, Genesis Care/21st Century Oncology, Fort Walton Beach, Florida, USA.
Zitieren
APA
Hill, L. W., & Jack, D. (2024). A slope method for the determination of electron energy for quality assurance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14369
Vancouver
Hill LW, Jack D. A slope method for the determination of electron energy for quality assurance. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14369
BibTeX
@article{Hill2024,
author = {Hill, Larry W and Jack, David},
title = {A slope method for the determination of electron energy for quality assurance},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14369}
}
EUVIMED