Optimization of reconstruction in quantitative brain PET images: Benefits from PSF modeling and correction of edge artifacts

Emilie Verrecchia-Ramos, Merwan Ginet, Olivier Morel, Marc Engels-Deutsch, Sinan Ben Mahmoud, Paul Retif

Medical Physics · 2024

DOI: 10.1002/mp.17419 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Emilie Verrecchia-Ramos and 5 co-authors. Institution: CHR Metz-Thionville, Department of Medical Physics, Mercy Hospital, Ars-Laquenexy, France. Major MEDLINE topics: Artifacts, Phantoms, Imaging, Brain, Image Processing, Computer-Assisted. Cited 2 times so far (Crossref); 39 references.

AbstraktsSubjectsBibliogrāfiskie datiCiteRelated studies

Abstract (original language)

Background

Modern PET reconstruction algorithms incorporate point‐spread‐function (PSF) correction to mitigate partial volume effect. However, PSF correction can introduce edge artifacts that lead to quantification errors. Consequently, current international guidelines advise against using PSF correction in brain PET reconstruction.

Purpose

We aimed to investigate PSF‐induced quantification errors in recent digital PET systems and identify conditions that mitigate them. This study utilized brain PET imaging with alginate‐based realistic phantoms, simulating lesion‐to‐background activity ratios of 10:1 and 2:1, with eleven reconstruction parameter sets.

Methods

Phantoms were prepared using a commercial anthropomorphic head phantom and two homemade inserts. Each insert contained a homogeneous 18 F‐FDG alginate background with hot spheres of varying diameter (3, 4, 6, 8, 10, 12, and 15 mm). PET imaging was conducted on a digital PET‐CT system Biograph Vision 600 (Siemens), with a 10 min scan duration. Imaging was performed with and without PSF correction, with 2, 4, 6, 12, 18, or 24 iterations in reconstruction, and with or without additional Gaussian postfiltering. We assessed the recovery coefficient (RC), contrast recovery coefficient (CRC), variability, and CRC‐to‐variability ratios for each sphere size and reconstruction parameter set.

Results

PSF‐corrected images of the 10:1 spheres exhibited a nonmonotonic CRC‐to‐sphere diameter relationship due to edge artifacts overshoot in the 10 mm‐diameter sphere. In contrast, PSF images of the 2:1 spheres showed a monotonically increasing relationship. Non‐PSF images of both phantoms showed an expected monotonically increasing CRC‐to‐sphere diameter relationship but with lower CRC values compared to PSF images. The nonmonotonic relationship observed with 10:1 spheres was mitigated by applying a 3‐mm FWHM Gaussian postfiltering. For both phantoms, reconstructions with 6 iterations, PSF correction, and additional 3‐mm FWHM Gaussian postfiltering demonstrated the highest CRC‐to‐variability ratios.

Conclusions

Our findings indicate that Gaussian postfiltering suppresses PSF artifacts. This parameter set corrected the nonmonotonic CRC‐to‐sphere diameter relationship and improved the CRC‐to‐variability ratio compared to non‐PSF reconstructions. Therefore, to enhance lesion detectability without compromising quantification accuracy, PSF correction coupled with Gaussian postfiltering should be used in brain PET.

Subjects

MeSH terms (MEDLINE)

ArtifactsPhantoms, ImagingBrainImage Processing, Computer-AssistedPositron-Emission TomographyHumans

Author keywords

PSF correctionSUVquantification

Authors and institutions

  • CHR Metz-Thionville, Department of Medical Physics, Mercy Hospital, Ars-Laquenexy, France.
  • CHR Metz-Thionville, Department of Nuclear Medicine, Mercy Hospital, Ars-Laquenexy, France.
  • CHR Metz-Thionville, Department of Odontology, Mercy Hospital, Ars-Laquenexy, France.
  • CNRS, LEM3, Université de Lorraine, Nancy, France.
  • CNRS, CRAN, Université de Lorraine, Nancy, France.

Cite

APA

Verrecchia-Ramos, E., Ginet, M., Morel, O., Engels-Deutsch, M., Ben Mahmoud, S., & Retif, P. (2024). Optimization of reconstruction in quantitative brain PET images: Benefits from PSF modeling and correction of edge artifacts. Medical Physics. https://doi.org/10.1002/mp.17419

Vancouver

Verrecchia-Ramos E, Ginet M, Morel O, Engels-Deutsch M, Ben Mahmoud S, Retif P. Optimization of reconstruction in quantitative brain PET images: Benefits from PSF modeling and correction of edge artifacts. Medical Physics. 2024. doi:10.1002/mp.17419

BibTeX
@article{VerrecchiaRamos2024,
  author = {Verrecchia-Ramos, Emilie and Ginet, Merwan and Morel, Olivier and Engels-Deutsch, Marc and Ben Mahmoud, Sinan and Retif, Paul},
  title = {Optimization of reconstruction in quantitative brain PET images: Benefits from PSF modeling and correction of edge artifacts},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17419}
}
Start for free →