EUVIMEDObjevte EUVIMED
StudienprotokollOpen access · CC BY 4.0

AAPM WGTG51 Report 385: Addendum to the AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy electron beams

Bryan Muir, Stephen Davis, Sandeep Dhanesar, Yair Hillman, Viktor Iakovenko, Grace Gwe-Ya Kim, Victor Gabriel Leandro Alves, Yu Lei, Jessica Lowenstein, James Renaud, Arman Sarfehnia, Jeffrey Siebers and 1 more

Medical Physics · 2024

DOI: 10.1002/mp.17277 ↗

At a glance

Studienprotokoll in Medical Physics, 2024. Written by Bryan Muir and 12 co-authors. Institution: Metrology Research Centre, National Research Council of Canada, Ottawa, Ontario, Canada. Major MEDLINE topics: Electrons, Radiometry. Cited 10 times so far (Crossref); 69 references. The full text is freely available (open access, CC BY 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

An Addendum to the AAPM's TG‐51 protocol for the determination of absorbed dose to water is presented for electron beams with energies between 4 MeV and 22 MeV (). This updated formalism allows simplified calibration procedures, including the use of calibrated cylindrical ionization chambers in all electron beams without the use of a gradient correction. New data are provided for electron beams based on Monte Carlo simulations. Implementation guidance is provided. Components of the uncertainty budget in determining absorbed dose to water at the reference depth are discussed. Specifications for a reference‐class chamber in electron beams include chamber stability, settling, ion recombination behavior, and polarity dependence. Progress in electron beam reference dosimetry is reviewed. Although this report introduces some major changes (e.g., gradient corrections are implicitly included in the electron beam quality conversion factors), they serve to simplify the calibration procedure. Results for absorbed dose per linac monitor unit are expected to be up to approximately 2 % higher using this Addendum compared to using the original TG‐51 protocol.

Subjects

MeSH terms (MEDLINE)

ElectronsRadiometryCalibrationMonte Carlo MethodHumansUncertaintyReference StandardsWaterRadiotherapy Dosage

Author keywords

k Q $k_{Q}$absorbed‐dose calibration coefficientbeam quality conversion factordosimetry protocolelectron beamionization chamberuncertainty analysis

Substances

Water

Authors and institutions

  • Metrology Research Centre, National Research Council of Canada, Ottawa, Ontario, Canada.
  • Department of Radiation Oncology, Miami Cancer Institute, Miami, Florida, USA.
  • Department of Radiation Oncology, Houston Methodist Hospital, Houston, Texa, USA.
  • Department of Radiation Oncology, Sharett Institute of Oncology, Hadassah Medical Center, Hebrew University of Jerusalem, Jerusalem, Israel.
  • UT Southwestern Medical Center, Dallas, Texa, USA.
  • Department of Radiation Medicine and Applied Sciences, UC San Diego School of Medicine, San Diego, California, USA.

Cite

APA

Muir, B., Davis, S., Dhanesar, S., Hillman, Y., Iakovenko, V., Kim, G. G., Alves, V. G. L., Lei, Y., Lowenstein, J., Renaud, J., Sarfehnia, A., Siebers, J., & Tantôt, L. (2024). AAPM WGTG51 Report 385: Addendum to the AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy electron beams. Medical Physics. https://doi.org/10.1002/mp.17277

Vancouver

Muir B, Davis S, Dhanesar S, Hillman Y, Iakovenko V, Kim GG, et al. AAPM WGTG51 Report 385: Addendum to the AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy electron beams. Medical Physics. 2024. doi:10.1002/mp.17277

BibTeX
@article{Muir2024,
  author = {Muir, Bryan and Davis, Stephen and Dhanesar, Sandeep and Hillman, Yair and Iakovenko, Viktor and Kim, Grace Gwe-Ya and Alves, Victor Gabriel Leandro and Lei, Yu and Lowenstein, Jessica and Renaud, James and Sarfehnia, Arman and Siebers, Jeffrey and Tantôt, Laurent},
  title = {AAPM WGTG51 Report 385: Addendum to the AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy electron beams},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17277}
}
Start for free →