Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities
Medical Physics · 2024
DOI: 10.1002/mp.17345 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Peter G F Watson und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Unit, McGill University, Montreal, Quebec, Canada. Hauptthemen laut MEDLINE: Radiometry, Monte Carlo Method. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The formalism proposed by Watson et al. allows users of the INTRABEAM (Carl Zeiss Medical AG, Jena, Germany) electronic brachytherapy system to accurately determine the absorbed dose to water, in the absence of a primary dosimetry standard. However, all published values are for PTW 34013 ionization chambers calibrated in a TW30 reference beam, traceable to PTB (Germany). For North American users, it would be advantageous to have data for chambers calibrated in a kV reference beam maintained by the National Institute of Standards and Technology (NIST).
Purpose
In this work, we determine for a PTW 34013 chamber calibrated in three NIST‐traceable reference beams: M30, L40, and L50.
Methods
Using available photon spectra data for M30, L40, and L50 reference beam qualities, Monte Carlo simulations using EGSnrc were performed to calculate the ratio of the absorbed dose to the PTW 34013 chamber air cavity to air‐kerma () for these beams. From this ratio, as a function of depth in water was determined. The effect of the use of a buildup foil was also investigated. An uncertainty analysis considering both the Type A and Type B uncertainties in the calculation of was performed.
Results
The largest difference in was found between L50 and TW30, with a relative decrease of 1.4% (no buildup) to 1.6% (buildup). For M30 and L40, the differences were minimal compared with measurement uncertainties.
Conclusions
We report values for three NIST‐traceable kV reference beams. This study reinforces the feasibility of adapting the Watson et al. methodology using different kV reference beams, facilitating the use of INTRABEAM in North America and ensuring the continuity and accuracy of dosimetry standards in intraoperative radiation therapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics Unit, McGill University, Montreal, Quebec, Canada.
- Department of Radiation Oncology, Miami Cancer Institute, Miami, Florida, USA.
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Zitieren
APA
Watson, P. G. F., Davis, S., & Culberson, W. S. (2024). Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities. Medical Physics. https://doi.org/10.1002/mp.17345
Vancouver
Watson PGF, Davis S, Culberson WS. Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities. Medical Physics. 2024. doi:10.1002/mp.17345
BibTeX
@article{Watson2024,
author = {Watson, Peter G F and Davis, Stephen and Culberson, Wesley S},
title = {Technical note: Determination of CQ$C_Q$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17345}
}
EUVIMED