EUVIMEDScopri EUVIMED
Crossref

Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy

Gustavo H Píriz, Gabriel A Gonzalez-Sprinberg, Facundo Ballester, Javier Vijande

Medical Physics · 2024

DOI: 10.1002/mp.17035 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Gustavo H Píriz and 3 co-authors. Institution: Medical Physics Unit, Faculty of Sciences, University of the Republic, Montevideo, Uruguay. Major MEDLINE topics: Brachytherapy, Cobalt Radioisotopes, Radiometry, Monte Carlo Method. 21 references. Funded by: ''Ministerio de Ciencia e Innovación″ (MCIN), ''Generalitat Valenciana″ (GVA).

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Non‐melanoma skin cancer is one of the most common types of cancer and one of the main approaches is brachytherapy. For small lesions, the treatment of this cancer with brachytherapy can be done with two commercial applicators, one of these is the Large Field Valencia Applicators (LFVA).

Purpose

The aim of this study is to test the capabilities of the LFVA to use clinically 60 Co sources instead of the 192 Ir ones. This study was designed for the same dwell positions and weights for both sources.

Methods

The Penelope Monte Carlo code was used to evaluate dose distribution in a water phantom when a 60 Co source is considered. The LFVA design and the optimized dwell weights reported for the case of 192 Ir are maintained with the only exception of the dwell weight of the central position, that was increased. 2D dose distributions, field flatness, symmetry and the leakage dose distribution around the applicator were calculated.

Results

When comparing the dose distributions of both sources, field flatness and symmetry remain unchanged. The only evident difference is an increase of the penumbra regions for all depths when using the 60 Co source. Regarding leakage, the maximum dose within the air volume surrounding the applicator is in the order of 20% of the prescription dose for the 60 Co source, but it decreases to less than 5% at about 1 cm distance.

Conclusions

Flatness and symmetry remains unaltered as compared with 192 Ir sources, while an increase in leakage has been observed. This proves the feasibility of using the LFVA in a larger range of clinical applications.

Subjects

MeSH terms (MEDLINE)

BrachytherapyCobalt RadioisotopesRadiometryMonte Carlo MethodRadiotherapy DosagePhantoms, ImagingIridium RadioisotopesHumans

Author keywords

Monte CarloValencia applicatorsdosimetryskin brachytherapy

Substances

Cobalt RadioisotopesCobalt-60Iridium RadioisotopesIridium-192

Authors and institutions

  • Medical Physics Unit, Faculty of Sciences, University of the Republic, Montevideo, Uruguay.
  • Departamento de Física Atómica, Molecular y Nuclear, Universitat de Valencia (UV), Burjassot, Spain.
  • Unidad Mixta de Investigación en Radiofísica e Instrumentación Nuclear en Medicina (IRIMED), Ins-tituto de Investigación Sanitaria La Fe (IIS-La Fe), Universitat de Valencia (UV), València, Spain.
  • Instituto de Física Corpuscular, IFIC (UV-CSIC), Burjassot, Spain.

Funding

  • ''Ministerio de Ciencia e Innovación″ (MCIN)
  • ''Generalitat Valenciana″ (GVA)

Cite

APA

Píriz, G. H., Gonzalez-Sprinberg, G. A., Ballester, F., & Vijande, J. (2024). Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy. Medical Physics. https://doi.org/10.1002/mp.17035

Vancouver

Píriz GH, Gonzalez-Sprinberg GA, Ballester F, Vijande J. Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy. Medical Physics. 2024. doi:10.1002/mp.17035

BibTeX
@article{Priz2024,
  author = {Píriz, Gustavo H and Gonzalez-Sprinberg, Gabriel A and Ballester, Facundo and Vijande, Javier},
  title = {Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17035}
}
Start for free →