Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy
Medical Physics · 2024
DOI: 10.1002/mp.17035 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Gustavo H Píriz und 3 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Unit, Faculty of Sciences, University of the Republic, Montevideo, Uruguay. Hauptthemen laut MEDLINE: Brachytherapy, Cobalt Radioisotopes, Radiometry, Monte Carlo Method. Die Arbeit nennt 21 Literaturangaben. Gefördert von: ''Ministerio de Ciencia e Innovación″ (MCIN), ''Generalitat Valenciana″ (GVA).
Abstract (Originalsprache)
Background
Non‐melanoma skin cancer is one of the most common types of cancer and one of the main approaches is brachytherapy. For small lesions, the treatment of this cancer with brachytherapy can be done with two commercial applicators, one of these is the Large Field Valencia Applicators (LFVA).
Purpose
The aim of this study is to test the capabilities of the LFVA to use clinically 60 Co sources instead of the 192 Ir ones. This study was designed for the same dwell positions and weights for both sources.
Methods
The Penelope Monte Carlo code was used to evaluate dose distribution in a water phantom when a 60 Co source is considered. The LFVA design and the optimized dwell weights reported for the case of 192 Ir are maintained with the only exception of the dwell weight of the central position, that was increased. 2D dose distributions, field flatness, symmetry and the leakage dose distribution around the applicator were calculated.
Results
When comparing the dose distributions of both sources, field flatness and symmetry remain unchanged. The only evident difference is an increase of the penumbra regions for all depths when using the 60 Co source. Regarding leakage, the maximum dose within the air volume surrounding the applicator is in the order of 20% of the prescription dose for the 60 Co source, but it decreases to less than 5% at about 1 cm distance.
Conclusions
Flatness and symmetry remains unaltered as compared with 192 Ir sources, while an increase in leakage has been observed. This proves the feasibility of using the LFVA in a larger range of clinical applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Medical Physics Unit, Faculty of Sciences, University of the Republic, Montevideo, Uruguay.
- Departamento de Física Atómica, Molecular y Nuclear, Universitat de Valencia (UV), Burjassot, Spain.
- Unidad Mixta de Investigación en Radiofísica e Instrumentación Nuclear en Medicina (IRIMED), Ins-tituto de Investigación Sanitaria La Fe (IIS-La Fe), Universitat de Valencia (UV), València, Spain.
- Instituto de Física Corpuscular, IFIC (UV-CSIC), Burjassot, Spain.
Förderung
- ''Ministerio de Ciencia e Innovación″ (MCIN)
- ''Generalitat Valenciana″ (GVA)
Zitieren
APA
Píriz, G. H., Gonzalez-Sprinberg, G. A., Ballester, F., & Vijande, J. (2024). Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy. Medical Physics. https://doi.org/10.1002/mp.17035
Vancouver
Píriz GH, Gonzalez-Sprinberg GA, Ballester F, Vijande J. Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy. Medical Physics. 2024. doi:10.1002/mp.17035
BibTeX
@article{Priz2024,
author = {Píriz, Gustavo H and Gonzalez-Sprinberg, Gabriel A and Ballester, Facundo and Vijande, Javier},
title = {Dosimetry of Large Field Valencia applicators for Cobalt‐60‐based brachytherapy},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17035}
}
EUVIMED