Technical note: Cryostat transmission characterization for MR linac – temporal stability, clinical impact and change implementation

Urszula Jelen iD, Claire Pagulayan, Zoë Moutrie, Jason Arts, Armia George, Michael G Jameson

Medical Physics · 2024

DOI: 10.1002/mp.17021 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Urszula Jelen und 5 weiteren Autor:innen. Beteiligte Einrichtung: GenesisCare, Alexandria, NSW, Australia. Hauptthemen laut MEDLINE: Particle Accelerators, Radiotherapy Planning, Computer-Assisted, Radiosurgery. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 12 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In the Unity MR linac (Elekta AB, Stockholm, Sweden), the radiation beam traverses the cryostat and the coil support structure. The resulting beam attenuation must be considered for output calibration and its variation with gantry angle must be characterized in the treatment planning system (TPS).

Purpose

The aim of this work was to investigate the impact of a change of the cryostat transmission characterization (CTC) curve, due to the helium level modification, on clinical treatment plan dosimetry and to report on the experience with the CTC curve update.

Methods

Twenty stereotactic body radiotherapy (SBRT) treatment plans: 10 prostate and 10 oligo‐metastatic cancer plans, prepared with a beam model incorporating the CTC curve acquired at installation time, were re‐calculated using the model implementing CTC curve post helium top‐up. To account for the CTC change as well as to align our system to the recent reference conditions recommendations, the new model was commissioned with the emphasis on the specifics associated with the treatment plan adaptation and the existence of the offline and online TPS components.

Results

Average CTV mean dose reduction by 0.45% in prostate cases and average GTV mean dose reduction by 0.22% in oligo‐metastatic cases was observed. Updated model validation showcased good agreement between measurements and TPS calculations.

Conclusions

The agreement between CTC measurements demonstrates its temporal constancy and robustness of the measurement method employed. A helium fill level change was shown to affect the CTC and led to a small but systematic dose calculation inaccuracy. Finally, model validation and end‐to‐end testing results presented, underscore the minimal impact of transitioning to the new beam model and new reference conditions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Particle AcceleratorsRadiotherapy Planning, Computer-AssistedRadiosurgeryHumansMagnetic Resonance ImagingProstatic NeoplasmsMaleTime FactorsRadiotherapy DosageHeliumRadiometry

Stichwörter der Autor:innen

MRLMR‐linaccryostat characterizationdosimetry

Substanzen

Helium

Autor:innen und Institutionen

  • GenesisCare, Alexandria, NSW, Australia.
  • South Western Sydney Cancer Services, New South Wales Health, Liverpool, NSW, Australia.
  • South West Sydney Clinical Campuses, University of New South Wales, Warwick Farm, NSW, Australia.
  • Ingham Institute for Applied Medical Research, Liverpool, NSW, Australia.
  • Elekta Pty Ltd, North Sydney, NSW, Australia.
  • University of New South Wales, Sydney, NSW, Australia.

ORCID-Profile: Urszula Jelen

Zitieren

APA

Jelen, U., Pagulayan, C., Moutrie, Z., Arts, J., George, A., & Jameson, M. G. (2024). Technical note: Cryostat transmission characterization for MR linac – temporal stability, clinical impact and change implementation. Medical Physics. https://doi.org/10.1002/mp.17021

Vancouver

Jelen U, Pagulayan C, Moutrie Z, Arts J, George A, Jameson MG. Technical note: Cryostat transmission characterization for MR linac – temporal stability, clinical impact and change implementation. Medical Physics. 2024. doi:10.1002/mp.17021

BibTeX
@article{Jelen2024,
  author = {Jelen, Urszula and Pagulayan, Claire and Moutrie, Zoë and Arts, Jason and George, Armia and Jameson, Michael G},
  title = {Technical note: Cryostat transmission characterization for MR linac – temporal stability, clinical impact and change implementation},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17021}
}
Kostenlos starten →