MV‐based relative electron density estimation (iMREDe) for MR‐LINAC dose calculation

Marios Myronakis, Yue-Houng Hu, Matthew William Jacobson, Christopher Leigh Williams, Ross Isaac Berbeco

Medical Physics · 2024-03 · 51 (3) · 2155-2163

DOI: 10.1002/mp.16969 ↗

At a glance

Journal article in Medical Physics, published online on 2024-02-03. Written by Marios Myronakis and 4 co-authors. Institution: Department of Radiation Oncology, Brigham and Women's Hospital, Dana Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts, USA. Major MEDLINE topics: Electrons, Head and Neck Neoplasms. Cited 2 times so far (Crossref); 22 references. Funded by: ViewRay, Inc.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

MR‐LINAC systems have been increasingly utilized for real‐time imaging in adaptive treatments worldwide. Challenges in MR representation of air cavities and subsequent estimation of electron density maps impede planning efficiency and may lead to dose calculation uncertainties.

Purpose

To demonstrate the generation of accurate electron density maps using the primary MV beam with a flat‐panel imager.

Methods

The ViewRay MRIdian MR‐LINAC system was modeled digitally for Monte Carlo simulations. Iron shimming, the magnetic field, and the proposed flat panel detector were included in the model. The effect of the magnetic field on the detector response was investigated. Acquisition of projections over 360 degrees was simulated for digital phantoms of the Catphan 505 phantom and a patient treated for Head and Neck cancer. Shim patterns on the projections were removed and detector noise linearity was assessed. Electron density maps were generated for the digital patient phantom using the flat‐panel detector and compared with actual treatment planning CT generated electron density maps of the same patient.

Results

The effect of the magnetic field on the detector point‐spread function (PSF) was found to be substantial for field strengths above 50 mT. Shims correction in the projection images using air normalization and in‐painting effectively removed reconstruction artifacts without affecting noise linearity. The relative difference between reconstructed electron density maps from the proposed method and electron density maps generated from the treatment planning CT was 11% on average along all slices included in the iMREDe reconstruction.

Conclusions

The proposed iMREDe technique demonstrated the feasibility of generating accurate electron densities for the ViewRay MRIdian MR‐LINAC system with a flat‐panel imager and the primary MV beam. This work is a step towards reducing the time and effort required for adaptive radiotherapy in the current ViewRay MR‐LINAC systems.

Subjects

MeSH terms (MEDLINE)

ElectronsHead and Neck NeoplasmsHumansCone-Beam Computed TomographyPhantoms, ImagingMonte Carlo MethodParticle Accelerators

Author keywords

CBCTEPIDMR-LINACMV imagingelectron density

Authors and institutions

  • Department of Radiation Oncology, Brigham and Women's Hospital, Dana Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts, USA.

Funding

  • ViewRay, Inc

Cite

APA

Myronakis, M., Hu, Y. H., Jacobson, M. W., Williams, C. L., & Berbeco, R. I. (2024). MV‐based relative electron density estimation (iMREDe) for MR‐LINAC dose calculation. Medical Physics, 51(3), 2155-2163. https://doi.org/10.1002/mp.16969

Vancouver

Myronakis M, Hu YH, Jacobson MW, Williams CL, Berbeco RI. MV‐based relative electron density estimation (iMREDe) for MR‐LINAC dose calculation. Medical Physics. 2024;51(3):2155-2163. doi:10.1002/mp.16969

BibTeX
@article{Myronakis2024,
  author = {Myronakis, Marios and Hu, Yue-Houng and Jacobson, Matthew William and Williams, Christopher Leigh and Berbeco, Ross Isaac},
  title = {MV‐based relative electron density estimation (iMREDe) for MR‐LINAC dose calculation},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {3},
  pages = {2155--2163},
  doi = {10.1002/mp.16969}
}
Start for free →