Monte Carlo analysis of energy deposition and X‐ray fluence in cylindrical anode systems

M K M Alharbi, Abel Zhou, Rob Davidson iD, Mark Naunton, Chandra Makanjee

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70262 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von M K M Alharbi und 4 weiteren Autor:innen. Beteiligte Einrichtung: University of Canberra, Bruce ACT, Australia. Hauptthemen laut MEDLINE: Monte Carlo Method, Electrons, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Cylindrical anode X‐ray systems are increasingly used in multisource imaging; however, electron beam interactions with the curved anodes cause geometric distortions that alter energy deposition and X‐ray emission. Understanding these effects is key to optimizing system performance.

Purpose

This study uses Monte Carlo (MC) simulations to examine how electron beam size, anode radius, and polar angle influence energy deposition and X‐ray fluence in cylindrical anode setups, and to quantify the distortions and energy redistribution for improved X‐ray generation efficiency and beam stability.

Methods

MC simulations were performed with electron beams (0.5 × 0.5 mm and 2 × 2 mm, 120 keV) on cylindrical tungsten anodes with radii from 1 to 5 cm and polar angles from 20.5° to 71.8°. Energy deposition profiles, dimensions, and photon fluence distributions were analyzed using the FLUKA MC package, with mapping in cylindrical coordinates ( r ‐ ϕ ‐ z ).

Results

Energy deposition profiles varied with beam position and anode curvature. The axial full width at half maximum (FWHM) increased by up to 650% at larger polar angles, while the azimuthal FWHM decreased up to 50%. Larger anode radii reduced the azimuthal FWHM by up to 78%, with minimal changes in axial and radial components. Narrower beams (0.5 × 0.5 mm 2 ) produced smaller, more symmetric energy deposition profiles on anode surface. Overshoot occurred at small radii and large polar angles, leading to incomplete energy deposition.

Conclusions

Larger anode radii and moderate polar angles minimized energy deposition profiles distortion and improved X‐ray fluence uniformity and production efficiency. Overshooting at small radii and large angles caused deformation, emphasizing the need for precise beam positioning to balance distortion, uniformity, and efficiency. These results define the geometric limits for effective energy deposition profile formation in cylindrical anode systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodElectronsPhantoms, ImagingRadiotherapy Planning, Computer-AssistedParticle AcceleratorsElectrodesX-RaysHumansPhotonsComputer Simulation

Stichwörter der Autor:innen

Monte Carlo simulationX‐ray fluenceX‐ray tubecylindrical anodefocal spot distortion

Autor:innen und Institutionen

  • University of Canberra, Bruce ACT, Australia.
  • Taif University, Taif, Saudi Arabia.
  • Singapore Institute of Technology, Singapore, Singapore.
  • College of Medicine, Nursing & Health Sciences, Fiji National University, Suva, Fiji.

ORCID-Profile: Rob Davidson

Zitieren

APA

Alharbi, M. K. M., Zhou, A., Davidson, R., Naunton, M., & Makanjee, C. (2025). Monte Carlo analysis of energy deposition and X‐ray fluence in cylindrical anode systems. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70262

Vancouver

Alharbi MKM, Zhou A, Davidson R, Naunton M, Makanjee C. Monte Carlo analysis of energy deposition and X‐ray fluence in cylindrical anode systems. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70262

BibTeX
@article{Alharbi2025,
  author = {Alharbi, M K M and Zhou, Abel and Davidson, Rob and Naunton, Mark and Makanjee, Chandra},
  title = {Monte Carlo analysis of energy deposition and X‐ray fluence in cylindrical anode systems},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70262}
}
Kostenlos starten →