Streamlined pin‐ridge‐filter design for single‐energy proton FLASH planning
Medical Physics · 2024-04 · 51 (4) · 2955-2966
DOI: 10.1002/mp.16939 ↗At a glance
Journal article in Medical Physics, published online on 2024-01-12. Written by Chaoqiong Ma and 7 co-authors. Institution: Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, Georgia, USA. Major MEDLINE topics: Neoplasms, Radiotherapy, Intensity-Modulated, Proton Therapy. Cited 21 times so far (Crossref); 41 references. Funded by: National Cancer Institute of the National Institutes of Health.
Abstract (original language)
Background
FLASH radiotherapy (FLASH‐RT) with ultra‐high dose rate has yielded promising results in reducing normal tissue toxicity while maintaining tumor control. Planning with single‐energy proton beams modulated by ridge filters (RFs) has been demonstrated feasible for FLASH‐RT.
Purpose
This study explored the feasibility of a streamlined pin‐shaped RF (pin‐RF) design, characterized by coarse resolution and sparsely distributed ridge pins, for single‐energy proton FLASH planning.
Methods
An inverse planning framework integrated within a treatment planning system was established to design streamlined pin RFs for single‐energy FLASH planning. The framework involves generating a multi‐energy proton beam plan using intensity‐modulated proton therapy (IMPT) planning based on downstream energy modulation strategy (IMPT‐DS), followed by a nested pencil‐beam‐direction‐based (PBD‐based) spot reduction process to iteratively reduce the total number of PBDs and energy layers along each PBD for the IMPT‐DS plan. The IMPT‐DS plan is then translated into the pin‐RFs and the single‐energy beam configurations for IMPT planning with pin‐RFs (IMPT‐RF). This framework was validated on three lung cases, quantifying the FLASH dose of the IMPT‐RF plan using the FLASH effectiveness model. The FLASH dose was then compared to the reference dose of a conventional IMPT plan to measure the clinical benefit of the FLASH planning technique.
Results
The IMPT‐RF plans closely matched the corresponding IMPT‐DS plans in high dose conformity (conformity index of
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, Georgia, USA.
Funding
- National Cancer Institute of the National Institutes of Health
Cite
APA
Ma, C., Zhou, J., Chang, C. W., Wang, Y., Patel, P. R., Yu, D. S., Tian, S., & Yang, X. (2024). Streamlined pin‐ridge‐filter design for single‐energy proton FLASH planning. Medical Physics, 51(4), 2955-2966. https://doi.org/10.1002/mp.16939
Vancouver
Ma C, Zhou J, Chang CW, Wang Y, Patel PR, Yu DS, et al. Streamlined pin‐ridge‐filter design for single‐energy proton FLASH planning. Medical Physics. 2024;51(4):2955-2966. doi:10.1002/mp.16939
BibTeX
@article{Ma2024,
author = {Ma, Chaoqiong and Zhou, Jun and Chang, Chih-Wei and Wang, Yinan and Patel, Pretesh R and Yu, David S and Tian, Sibo and Yang, Xiaofeng},
title = {Streamlined pin‐ridge‐filter design for single‐energy proton FLASH planning},
journal = {Medical Physics},
year = {2024},
volume = {51},
number = {4},
pages = {2955--2966},
doi = {10.1002/mp.16939}
}
EUVIMED