Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons
Medical Physics · 2024
DOI: 10.1002/mp.16925 ↗At a glance
Journal article in Medical Physics, 2024. Written by Ashley J Cetnar and 3 co-authors. Institution: Department of Radiation Oncology, James Cancer Hospital and Solove Research Institute, The Ohio State University, Columbus, Ohio, USA. Major MEDLINE topics: Electrons, Particle Accelerators. Cited 11 times so far (Crossref); 37 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: Varian Medical Systems.
Abstract (original language)
Background
Ultra‐high dose rate radiation (UHDR) is being explored by researchers in promise of advancing radiation therapy treatments.
Purpose
This work presents the commissioning of Varian's Flash Extension for research (FLEX) conversion of a Clinac to deliver UHDR electrons.
Methods
A Varian Clinac iX with the FLEX conversion was commissioned for non‐clinical research use with 16 MeV UHDR (16H) energy. This involved addition of new hardware, optimizing the electron gun voltages, radiofrequency (RF) power, and steering coils in order to maximize the accelerated electron beam current, sending the beam through custom scattering foils to produce the UHDR with 16H beam. Profiles and percent depth dose (PDD) measurements for 16H were obtained using radiochromic film in a custom vertical film holder and were compared to 16 MeV conventional electrons (16C). Dose rate and dose per pulse (DPP) were calculated from measured dose in film. Linearity and stability were assessed using an Advanced Markus ionization chamber.
Results
Energies for 16H and 16C had similar beam quality based on PDD measurements. Measurements at the head of the machine (61.3 cm SSD) with jaws set to 10×10 cm 2 showed the FWHM of the profile as 7.2 cm, with 3.4 Gy as the maximum DPP and instantaneous dose rate of 8.1E5 Gy/s. Measurements at 100 cm SSD with 10 cm standard cone showed the full width at half max (FWHM) of the profile as 10.5 cm, 1.08 Gy as the maximum DPP and instantaneous dose rate of 2.E5 Gy/s. Machine output with number of pulses was linear ( R = 1) from 1 to 99 delivered pulses. Output stability was measured within ±1% within the same session and within ±2% for daily variations.
Conclusions
The FLEX conversion of the Clinac is able to generate UHDR electron beams which are reproducible with beam properties similar to clinically used electrons at 16 MeV. Having a platform which can quickly transition between UHDR and conventional modes (
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Radiation Oncology, James Cancer Hospital and Solove Research Institute, The Ohio State University, Columbus, Ohio, USA.
Funding
- Varian Medical Systems
Cite
APA
Cetnar, A. J., Jain, S., Gupta, N., & Chakravarti, A. (2024). Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons. Medical Physics. https://doi.org/10.1002/mp.16925
Vancouver
Cetnar AJ, Jain S, Gupta N, Chakravarti A. Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons. Medical Physics. 2024. doi:10.1002/mp.16925
BibTeX
@article{Cetnar2024,
author = {Cetnar, Ashley J and Jain, Sagarika and Gupta, Nilendu and Chakravarti, Arnab},
title = {Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16925}
}
EUVIMED